beautypg.com

Nova-Tech pH 110 User Manual

Page 18

background image

Instruction Manual

CyberScan pH 11 / 110

14

7. Rinse the probe with de-ionised water or a rinse

solution, and place it in the next pH buffer.

8. Follow steps 5 and 6 for additional calibration

points. See Figure 12.

9. When all the calibration points set in the Unit

Configuration Setup (see P1.3 Select Number
of pH Calibration Points on page 36) are
completed, the meter returns to Measurement
mode automatically. However, if you wish to
terminate the calibration without completing the
number of points as set in the Unit
Configuration Setup menu, press CAL/MEAS to
return to pH measurement mode.

NOTES:

1. To exit from pH calibration mode without confirming

calibration, DO NOT press ENTER in step 6. Press
CAL/MEAS instead.

2. To limit the number of pH buffer values available during

calibration, see P1.3 Select Number of pH Calibration Points
on page 36.

CAL

pH

pH

READY

4.15

4.01

CAL

pH

pH

READY

4.01

4.01

ENTER

HOLD

Figure 12: Proceed to next

calibration point

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682

This manual is related to the following products: