beautypg.com

P 1.3 – Nova-Tech pH 110 User Manual

Page 40

background image

Instruction Manual

CyberScan pH 11 / 110

36

9.2.3 P1.3 Select Number of pH Calibration Points

Program P1.3 lets you select the number of calibration points that appear in pH
calibration mode: 2, 3, 4, or 5 (or 6 with DIN buffer option). This lets the meter scroll
through the calibration points more quickly if you regularly calibrate at less than 5
points.

From measurement mode:

1.

Press SETUP key to enter Set Up mode.

2. Press the MI/Ÿ or MR/ź keys to scroll through

subgroups until you view parameter P1.0.

3. Press the HOLD/ENTER key three times to

select parameter 1.3.

4. Press the MI/Ÿ or MR/ź keys to select 2, 3, 4,

or 5 point pH calibration (up to 6 points for DIN
calibration set with pH 110 meter).

5. Press the HOLD/ENTER key to confirm

selection and to return to the subgroup menu.
Press CAL/MEAS key to return to measurement
mode.

See Figure 32.

Figure 32: P1.3 - Select

number of calibration points

SETUP

3 p

p 1.3

SETUP

4 p

p 1.3

SETUP

5 p

p 1.3

SETUP

2 p

p 1.3

MR

MI

MR

MI

MR

MI

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682

This manual is related to the following products: