beautypg.com

Error messages, Specifications – Nova-Tech pH-Dissolved Oxygen Meter User Manual

Page 32

background image

63

62

11. Error Messages

LCD

Indicates

Cause

Solution

Display

Err

Unrecognized input

Wrong input in

Release key. Select

annunciator

from keypad

selected mode.

valid operations
depending on mode.

CAL & Err

Calibration error

Wrong value input

Check your input value,

annunciators

at calibration.

clean probe.

blink

Dirty probe.

See Calibration sections or
Probe Maintenance section.

Battery

Low battery level

Need new batteries

Clean battery contacts.

indicator blinks

or battery

Replace batteries with fresh

connection is bad.

ones, noting polarity.

* See “Warranty” and “Return of Items” on page 67

If an error message appears in the primary
display (the upper row of larger digits),
switching off the meter and switching it
on again may eliminate the error message.

If error persists, or the meter shows incorrect
values, return the meter.

For a complete diagram of the display,
see page 5.

12. Specifications

Mode

pH

mg/l (ppm)

% saturation

Temperature

Range

-2 to 16 pH

0.00 to

0.0 to 200.0%

0.0 to 50.0°C /

20.00 mg/l (ppm)

32.0 to 122.0°F

Resolution

0.01 pH

0.01 mg/l (ppm)

0.1%

0.1°C / 0.1°F

Accuracy

±0.01 pH

±1.5% full scale

±1.5% full scale

±0.3°C / ±0.5°F

Calibration

up to 5 points: one or two points (100% and 0% saturation);

offset up to

pH 1.68; 4.01;

plus separate mg/l (ppm) slope

±5°C or ±10°F

7.00; 10.01; 12.45

Temperature compensation: automatic from 0 to 50°C

Salinity correction:

Range: 0.0 to 50.0 ppt
Resolution: 0.1 ppt
Method: Key in manually and automatic correction

Barometric pressure correction:

Range: 500 to 1499 mm Hg / 66.6 to 199.9 kPa
Resolution: 1 mm Hg / 0.1 kPa
Method: Key in manually and automatic correction

DO Probe type: galvanic probe

Display: dual LCD

Backlit display: select from eight levels of brightness

Operating temperature: 0 to 50°C

Power: four 1.5 V AAA batteries (included)

Battery life: > 100 hours continuous use
use of backlit display decreases battery life

Dimensions:

Meter: 7.5"L x 3.75"W x 2.25"H (19 x 9.5 x 5.7 cm)
Boxed: 9.2"L x 9.2"W x 2.75"H (23 x 23 x 7 cm)
DO Probe: 6.8"L x 1.3" dia (173 mm L x 32 mm dia) with 10-ft cable
pH Electrode: 6.0"L x 1.0" dia (152 mm L x 25 mm dia) with 10-ft cable

Weight:

Meter: 1.0 lb (0.5 kg)
Boxed: 2.0 lbs (0.9 kg)

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682