Manual temperature compensation – Nova-Tech pH-Dissolved Oxygen Meter User Manual

Page 13

background image

24

25

Taking measurements with READY indicator selected on

If the READY indicator has been activated, the READY annunciator lights when
the reading is stable. Switch the READY indicator on or off in Set up program P3.1—
see page 35 for directions.

Taking measurements with the auto endpoint feature selected on

When a reading is stable for more than 5 seconds, the auto endpoint feature will
automatically “HOLD” the reading. The “HOLD” indicator appears on the left side
of the display. Press the HOLD key to release the reading. Switch the Auto endpoint
feature on or off in Set up program P3.1—see page 35 for directions.

IMPORTANT: For manual compensation,
you must disconnect the DO or tempera-
ture probe (see page 9).

1.

Switch the meter on. Press the MODE
key
to select pH mode.

2.

Press the CAL/MEAS key to enter pH
calibration mode. The CAL indicator will
appear above the primary display.

3.

While in pH calibration mode, press
the MODE key
to enter temperature
calibration mode. The primary display
shows the current temperature setting
and the secondary display shows the
default value 25°C.

See figure

4.

Check the temperature of your sample
using an accurate thermometer.

5.

Press the

or keys to offset the

temperature to the measured value
from step 4.

See figure

6.

Press ENTER to confirm the selected
temperature and to return to the pH
measurement mode.

The meter will now compensate pH
readings for the manually set temperature.

Notes

To exit this program without confirming the
manual temperature compensation value,
DO NOT press ENTER in step 6. Press
CAL/MEAS instead.

Manual Temperature Compensation