5 calibration setup, Calibration setup – Nova-Tech PGL Series User Manual
Page 34
1) Use the [Up] and [Down] keys to increase or decrease the value for setting.
Press [Enter] to accept the setting and advance to the next item in the menu
2) Press [Mode] to advance to setting of the next parameter or [Esc] to return to
normal weighing
12.5 CALIBRATION SETUP
This menu allows the user to set the calibration parameters.
Press [Enter] when “CAL SETUP” is displayed to select the calibration parameters
The options for each parameter can be scrolled through by using the [Up] or
[Down] key
ENABLE
NO =operator calibration is disabled
YES=operator calibration is enabled
CAL REPORT
On = Enabled. Prints out Calibration report after
successful calibration
OFF = Disabled
TIME CAL
On = Enabled. Select time from 1 to 24 hours.
OFF = Disabled
Press [Mode] to advance to setting of the next menu which is “PASSCODES” or
[Esc] to return to normal weighing.
34 |
P a g e
© Adam Equipment Company 2008