10.0  BATTERY

 

OPERATION

The scale may operate from a built in lead acid battery.   During operation with the 
battery the backlight should be disabled for best battery life. Typical battery life 20 
hours. 

If  possible  disable  the  backlight  automatically  if  the  external  power  supply  is  not 
turned on. 

11.0  ERROR

 

CHECKING 

During weighing the balance is constantly checking to see if the balance is operating 
within the limited parameters.  The errors likely to occur are: 

 

A/D counts below lowest allowed value 

 

A/D counts above highest allowed value 

 

A/D not operating 

 

Maximum capacity exceeded 

Other errors may be detected during special functions or operations.  These will be 
described in the section that applies. 

Error messages and the reasons are: 

Concerning calibration 
ERROR StAb 

Calibration could not be completed  

because the results were not stable  

ERROR LOW or  ERROR HIGH 

Calibration constant not within 20% of old calibration 
constant

Concerning weighing 
ERROR LOW 

Weight display is below zero by >4%max 

ERROR HIGH 

Weight is above maximum plus 90d 

ERROR ZERO 

Weight  on  pan  exceeds  zero  limit  when  power  is 
applied. 

29 |