7) The balance will request the weight in air by displaying “AIR WEIGHT”. Place the 

glass plumb supplied with the density determination kit in air on the weighing 
pan and press [Enter] to start the air weighing 

8) On  completion  of  the  air  weighing,  the  balance  will  request  the  weight  in 

liquid  by  displaying  “LIQUID WT”.  Submerge  the  glass  disk  in  the  liquid  and 
press the [Enter] key. The balance will compute the density of the liquid and 
display it. Remove the item from the pan 

9) Press  [Mode]  to  continue  with  a  new  sample  or  press  [Mode]  then  [Esc]  to 

return to normal weighing 

21 |

P a g e

© Adam Equipment Company 2008