9) The  balance  will  request  the  weight  of  the  sample  in  air  by  displaying  “AIR

WEIGHT”. Place the item on the pan or receptacle, if the  density kit  is  used. 
Press [Enter] to determine the value 

10) After completion of the air weighing, the balance will request the weight in 

liquid  by  displaying  “LIQUID WT”.  Submerge  the  item  in  the  liquid  and  press 
[Enter] to start the liquid weighing. The balance will compute the density of 
the sample and display it as “DENSITY XXXX g/cc”

11) After completion of the liquid weighing, remove the item from the pan 

12) Press  [Mode]  to  continue  with  a  new  sample    or  press    [Esc]  to  return  to 

normal weighing 

DENSITY OF LIQUID 

When  finding  the  density  of  a  liquid,  it  is  necessary  to  weigh  a  sample  of  known 
volume in air and then in the liquid. The volume of the sample must be entered by 
the user. The last known volume is stored for use at any time. 

If  using  the  density  determination  kit,  the  volume  of  the  plumb  is  marked  on  its 
support, i.e. 10.123 

Steps: 

1) Press [Mode] and then [Up] and [Down] to select Density mode. 

2) Press [Enter] to select the Density mode 

3) Use [Up] and [Down] to scroll through the solid or liquid method 

4) When “DENSITY LIqUId” is displayed, press [Enter] to enter the liquids method 

5) The  volume  will  be  asked  for.  Enter  the  volume  using  the  numeric  entry 

method (see section 5.1) or continue using the last volume entered 

6) Press [Enter] to continue 