7.0 FUNCTIONS
When weighing, the user can access the applications that have been enabled (see
section 12.2).
• Weighing
• Parts counting
• Percent weighing
• Check weighing
• Density determination
The selectable functions can be enabled using a similar method to the Units above
by turning the functions on or off.
7.1
PARTS COUNTING
This allows the user to weigh a sample of parts to compute an average unit weight
and then determine the number of items being weighed by dividing the net weight
by the unit weight value. The result is always a whole number of parts.
The balance has a preset number of parts to be used as a sample. These values are
10, 25, 50 or 100 items.
Steps:
1) Press [Mode] and then the [Up] or [Down] key to select parts counting,
“PARTS” will be displayed
2) Enter parts counting by pressing [Enter]
3) Press the [Up] or [Down] key to select the sample size, “REF QTY”, 10, 25, 50,
100, etc., then press [Enter] to confirm
4) When “LOAD XX Pcs” is shown place XX number of items on the pan and press
[Enter] to compute the average piece weight. Display will indicate the busy
symbol
5) Remove the sample when display shows “XX Pcs” and then place an unknown
quantity on the pan. The balance will then compute the number of parts
based upon the average piece weight. The display will show the result in Pcs
6) To count another item press [Mode] and continue from step 2.
15 |
P a g e
© Adam Equipment Company 2008
7) To return to normal weighing, press [Mode] then press [Esc]