Data Format for normal weighing operations, parts counting or recalling of totals
from memory will all be different.
Normal Output:
G S _ X X X
.
X X u u u
GS 123.45g GS for Gross Weight, NT for Net Weight and u for unit of weight
No. 1 This number increments every time a new value is stored in memory
Total 123.45g The total value stored in memory
Input command format:
The scale can be controlled with the following commands. The commands must be
sent in upper case letters, i.e. “T” not “t”.
T
Tares the scale to display the net weight. This is the same as pressing [Tare]
key.
Z
Sets the zero point for all subsequent weighing. Display shows zero.
T5.345
Would be same as entering a preset tare value of 5.345 from keypad
P
Prints the results to a PC or printer using the optional RS‐232 interface. It also
adds the value to the accumulation memory if the accumulation function is
not set to automatic.
25 |
P a g e