9.0 SPECIFICATIONS
9.1
TECHNICAL
SPECIFICATIONS
HCB123 HCB153 HCB302 HCB602 HCB602H
HCB1002 HCB1502 HCB3001
Maximum Capacity
120g
150g
300g
600g
600g
1000g
1500g
3000g
Readability
0.001g
0.005g
0.01g
0.02g
0.01g
0.01 g
0.05g
0.1g
Repeatability(s.d.)
0.002g
0.005g
0.01g
0.02g
0.01g
0.01 g
0.05g
0.1g
Linearity ±
0.04g
0.01g
0.02g
0.04g
0.02g
0.02 g
0.1g
0.2g
Pan
120mm / 4.7” ø
Draft shield
Included as standard
Units of Measure
g / ct /OZ / d / GN / OZt / dWt / MM /
TL.T / TL.C / TL.t /t / N /g2.
g / ct / Lb / OZ / d / GN / OZt / dWt / MM
/ TL.T / TL.C / TL.t /t / N /g2.
9.2
COMMON
SPECIFICATIONS
Interface
USB and RS‐232, bi‐directional
Stabilization Time
2 Seconds typical
Operating Temperature
0°C to 40°C / 32°F to 104°F
Power supply (external)
12VDC 800 mA
Calibration
HandiCal Internal calibration or external calibration ‐ User selectable
Display
18mm high 6 digits LCD
With auto backlight and loading bar graph
Scale Housing
ABS Plastic with Stainless Steel Pan
Overall Dimensions (wxdxh)
170 x 245 x 80mm / 6.7” x 9.6” X 3.1”
Net Weight
1.5 kg / 3.3 lb
9.3
RS‐232
/
USB
SPECIFICATIONS
Either the RS‐232 or USB interface can be used at one time. You cannot use both
interfaces at the same time. See parameters 6.3 Setting The Printing Parameters /
Accumulation section for full information on settings
The standard Interface parameters are:
Connection details are:
Generally a Null model cable is required
for connection to a computer or printers.
RS‐232 output of weighing data
ASCII code
4800 Baud
8 data bits
No Parity
Connector: 9 pin d‐subminiature socket
Pin 3 Output
Pin 2 Input
Pin 5 Signal Ground
24 |
P a g e
© Adam Equipment Company 2008