whenever  a stable reading is present. 
The balance has to return to zero 
before another reading is sent via the 
interface.  

 

Press the [Mode] key to change the setting. Press [Tare] to confirm and move 
to the next parameter. 

7) Setting the baud rate. This is the transmission speed for communication with 

printers  and  computers.  The  selected  rate  must  match  that  of  the  other 
device for communications to work.  

The following settings are available: 

b 600

b 1200

b 2400

b 4800

b 9600

 

Press the [Mode] key to change the setting. Press [Tare] to confirm and move 
to the next parameter. 

8) Parity settings. Parity is a communications check. There are 3 settings that the 

Highland can work with as follows:  

8 n 1

8 data bits, no parity

7 E 1

7 data bits, even parity

7 0 1

7 data bits, odd parity

Press the [Mode] key to change the setting. Press [Tare] to confirm and move 
to the next parameter. 

9) The  display  will  go  back  showing  F3 SEr.  Press  the  [Print]  key  to  return  to 

weighing or press [Mode] to go to the next function. 

17 | 

P a g e  

 

 

 

 

 

 

© Adam Equipment Company 2008

 