0 parameters, 1 enabling weighing units, Parameters – Nova-Tech IHS 2a User Manual
Page 14: Enabling weighing units
6.0 PARAMETERS
The balance has 7 parameters that can be set by the user.
FUNCTION
SECTION
DESCRIPTION
F1 UNT
See section 6.1
Sets the units to be used
g
/
ct / Lb / oz / d / GN / OZt / dWt / MM / TL.T / TL.C / TL.t /t
/ N /g2.
F2 EL
See section 6.2
Sets the backlight
EL on
: backlight always on
EL AU
: backlight automatically turns on when a key is
pressed
EL oFF
: backlight always off
F3 SEr
See section 6.3
Sets the print parameters
F4 oFF
See section 6.4
Sets the auto power‐off parameter
F5 IEC
See section 6.5
Internal or external calibration select
F6 CA
See section 6.6
Cal mass fine adjust
tECH
Technical parameters setting mode / factory setting
6.1
ENABLING
WEIGHING
UNITS
You can enable and disable the weighing units available to the user when they press
the [Mode] key as described in section 4.5 Weighing Units.
1) To set this parameter press the [Mode] key during self‐checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Press the [Tare] key to view the current settings of each unit.
4) Pressing the [Tare] key will cycle through the other units along with their
current settings. For example, if oFF is displayed with the weighing unit Carats,
the user will not be able to use this unit while weighing.
5) Pressing the [Mode] key will change the setting of a particular unit. For
example, to enable the weighing unit Carats, change the setting to on by
pressing the [Mode] key.
14 |
P a g e
© Adam Equipment Company 2008