4 auto power off, 5 selecting the internal or external calibration, Auto power off – Nova-Tech IHS 2a User Manual
Page 18: Selecting the internal or external calibration
![background image](/manuals/340058/18/background.png)
6.4
AUTO
POWER
OFF
The auto power off function helps conserve power, when using the internal
rechargeable battery or AC adapter. The Auto switch‐off time may be set up by the
user.
1) To set this parameter press the [Mode] key during self‐checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Mode] key until F4 oFF is displayed.
4) Press the [Tare] key to view the current settings.
5) Press [Mode] to change the settings (0, 5, 10, 20 and 30 minutes).
6) Press [Tare] to store a particular setting. The display will return to F4 oFF
7) When F4 oFF is displayed press the [Print] key to return to weighing or press
[Mode] to go to the next function.
6.5
SELECTING
THE
INTERNAL
OR
EXTERNAL
CALIBRATION
You can select if the internal mass or an external mass is to be used to calibrate the
balance.
1) To set this parameter press the [Mode] key during self‐checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Mode] key until F5 IEC is displayed.
4) Press the [Tare] key to view the current settings.
5) To change the settings press the [Mode] key to change from INt (internal
calibration) or E (external calibration)
6) Press [Tare] to store a particular setting. The display will return to F5 IEC
7) When F5 IEC is displayed press the [Print] key to return to weighing or press
[Mode] to go to the next function.
18 |
P a g e
© Adam Equipment Company 2008