6.6 

ADJUST

 

THE

 

VALUE

 

OF

 

THE

 

INTERNAL

 

MASS 

The  internal  mass  value  stored  in  memory  can  be  adjusted  to  more  closely  match 
the value of the user’s external mass.     

1) To  set  this  parameter  press  the  [Mode]  key  during  self‐checking  test  when 

turning on the balance. 

2) After a few seconds, the display will show the first function F1 UNT.  

3) Keep pressing the [Mode] key until F6 CA is displayed. 

4) Press the [Tare] key to view the current settings.  

5) The display will show the current value with the first digit flashing. To change 

the value  press [%] to shift the position, press [Mode]  to increase the value 
and press [Print] to decrease the value. The internal calibration weight should 
only change between 95.000 and 105.000 grams or 495.00 and 505.00 grams 
(depending upon the model).

 

 

6) Press [Tare] to store a particular setting. The display will return to F6 CA 

7) When F6 CA is displayed press the [Print] key to return to weighing or press 

[Mode] to go to the next function. 

8) Pressing [Mode] will show TECH.  This function is a manufacturing function for 

qualified  technicians  only  and  you  can  press  [Mode]  again  to  scroll  through 
parameter options, or press [Print] to return to normal weighing. 

 

Changing the value of the internal weight will affect the calibration when using the 
internal calibration. To check if your internal weight is correct you should only use 
high quality weights that are accurate to the readability of the balance.    