beautypg.com

0 error messages – Nova-Tech CBK Series User Manual

Page 31

background image

© Adam Equipment Company 2007

29



14.0 ERROR MESSAGES

During the initial power-on testing or during operation, the scale may show an
error message. The meaning of the error messages is described below.

If an error message is shown, repeat the step that caused the message. If
the error message is still shown then contact your dealer for support.

ERROR
CODE

DESCRIPTION

POSSIBLE CAUSES

Err 1

Time input Error

Invalid time entry such as “268970” for
the time format “

H-m-S

”.

Err 2

Date input Error

34

th

day of a month is an invalid entry.

Err 4

Initial Zero is greater than
allowed (4% of maximum
capacity) when power is turned
on or when the

[Zero/Enter]

key is pressed.

Weight on the pan when turning the
scale on.
Excessive weight on the pan when
zeroing the scale.
Platform is not installed.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.

Err 6

A/D count is not correct when
turning the scale on.

Load cell is damaged.
Electronics is damaged.

Err 7

Percent input error

Percent function is entered with no
reference mass on the pan.

Err 8

High limit input error

Low limit is set first, then the high limit is
set lower than the low limit and high
limit not equal to zero.

Err 9

Low limit input error

High limit is set first, then the low limit is
set higher than the high limit and low
limit not equal to zero.

FAIL H

or

FAIL L

Calibration error

Improper calibration (should be within
+10% of the factory calibration). The old
calibration data will be retained until the