SET-UP OPERATING MODE






SP1200 will alternate these 2 displays until “RESET” is pressed.

Correct product must be displayed in the upper left hand corner.







To change product, press

key.








Press “CLEAR/FEED” key.






Enter product number desired:







With correct product displayed, check in upper right hand corner of the display for a temperature reading or the
words “No T/C”. If “No T/C” is displayed, press the “TC TOGGLE” key to switch on the temperature
compensation.





**All prover runs are to be made using temperature compensation.
**If “+426H” is displayed, the temperature probe or cable is open.
**If “-270.9L” is displayed, the temperature probe or cable is shorted.

PROVING MODE

LIN
Gm 0

No T/C
GO PROVE

00:00P

00-Mar-00

Press any key

to continue

**Flo-Cal**

Press CLEAR to
change product

Enter new
product #

LIN
Gm 0

No T/C
GO PROVE

LIN
Gm 0

-198.8c
GO PROVE

PRODUCT SELECTOR

0 = LOX

1 = LIN

2 = LAR

3 = LCO2

4 = N20

5 = CCO2

6 = MAPP

7 = LPG