Liquid Controls SP1200 User Manual
Page 10

Enter the trailer totalizer reading noted at the start of the prover. Press “ENTER” key.
Enter the trailer totalizer reading noted at the end of the prover. Press “ENTER” key
Enter the type of Factor used to calibrate the trailer totalizer.
Press “0” if CF (Calibration Factor = 1/K) or press “1” if KF (K-Factor direct entry for calibration)
Press “ENTER” key
Enter the CF from the trailer totalizer.
Press “ENTER”
Enter the KF from the trailer totalizer.
Press “ENTER”
Examples ONLY!
***********************
Calibration information
10:19A 11-Mar-97
Meter : 44795
Plant : 815
Tractor : 456
Trailer : 789
LIN -198.8c
Delivry GPM: 100.
Calib total: 101
Delivry TCF: 1.012512
- - -
Truck start: 0
Truck stop : 100
Truck total: 100
Truck C/K : CF
Truck Factr: .1
- - -
Truck % err: -.990
Set new CF: .1010
Press “PRINT” key
Additional prove tickets may
be printed by pressing the
“PRINT” key.
***********************
Calibration information
10:19A 11-Mar-97
Meter : 44795
Plant : 815
Tractor : 456
Trailer : 789
LIN -198.8c
Delivry GPM: 100.
Calib total: 101
Delivry TCF: 1.012512
- - -
Truck start: 0
Truck stop : 100
Truck total: 100
Truck C/K : KF
Truck Factr: 1
- - -
Truck % err: -.990
Set new KF: .9901
***NOTE: If incorrect data was entered press “ENTER” & reinsert prove data. Dial the “New CF” or “New
KF” into the trailer totalizer. Press “RESET” key.
Enter CF?
Enter KF?
Set new CF:
XXXX
Set new KF:
XXXX
Enter cal type
0=CF / 1=KF?
End reading?
LIN
Gm 120
-198.8c
GO PROVE