beautypg.com

Ufield correction – Liquid Controls T350N-P User Manual

Page 7

background image

Sponsler, Inc.

T350N-P Truck Totalizer/Printer

pg. 7

Doc# MN-T350N-P

For higher resolution (reading in gallons and tenths, gallons and hundredths,
etc.) multiply by 10 for tenths; by one for hundredths. The number dialed into
dipswitch S5 corresponds to the number of digits to the right of the decimal point.

Example:

Flowmeter “K” Factor = 230

CF= .004348

Desired display = gallons and tenths

Multiply by = .04348

S4 = 1

Factor Switches = 4348; S3 = 4, S2 = 3, S1 = 4, S0 = 8

Decimal point = S5 = 1

To adjust the calibration factor to reflect the turbines actual response to the
operating conditions apply the following formula:

Field Correction


New Calibration Factor = Actual Total

Meter Total

X Calibration Factor


Example #2:

Actual Total = 50

Meter Total = 52

C.F. = .004348

50/52 X .004348 = New Calibration Factor

.9615 X .004348 = .0041806

Insert 4181 into S3 – S0 respectively


In the above example .9615 denotes that the meter is operating .385% fast and
multiplying the present calibration factor (.004348) by the ratio of Actual Total:
Meter Total (.9615) reduces the calibration factor 3.85%.

Assume that rather than gallons, liters are the desired engineering unit.

Calibration Factor – Change of Calibration Engineering Units


Example #3:

K-Factor = 230 pulses per gallon

Liters = 3.785 per gallon


3.785

230

= Calibration Factor


.01646 = Calibration Factor for display of liters