Laurel Electronics LAUREATE SERIES SCALE-WEIGHT METER User Manual
Page 17
17
OTHER KEYSTROKES FOR SCALE METER SETUP
If the MENU
key does not work, see Section 10 “Enabling & Locking Out Menu Items.”
Press Menu
Select Key
Press Digit
Select Key
Press Value Select
Key
_
dC U
DC Volts
__
0.2U
__
2.0U
_
20.0U
200.0U
600.0U
0.2, 2, 20, 200, 660V FS
_
dC A
DC Amps
__
2.0A
_
20.0A
_
200.0A
_
5.0A
0.2, 20, 200 mA, 5A FS
_
InPut
DC signal condi-
tioner board
_
rAtio
Strain gauge & ratio
__
0.2U
__
2.0U
_
20.0U
0.2, 2, 20V FS.
_
Strn
Strain or ratiometric
_
20.0
_
_
__
50.0
_ _
100.0
_
250.0
_
500.0
20, 50, 100, 250, 500 mV FS voltage
_
InPut
Load cell signal
conditioner board
_
dC u U
DC millivolts
_
20.0
_
_
__
50.0
_ _
100.0
_
250.0
_
500.0
20, 50, 100, 250, 500 mV FS voltage
00000
Dummy zero
0
No dummy zero (display to 99999)
1
Dummy zero (display to 999990)
00000
Power line frequency
0
Noise minimized for 60 Hz
1
Noise minimized for 50 Hz
00000
Peak Key Action
0
Display peak
1
Tare action
00000
Scaling method
0
Scale and offset method
1
Coordinates of 2 points method
SEtuP
Meter Setup
00000
Control inputs 1 & 2
or both:
True = logic 1 (0V or
tied to digital ground)
False = logic 0 (5V or
open)
Input 1
Input 2
Both 1 & 2
0
Meter Reset Meter Hold Meter Reset
1
Funct Reset Peak Displ
Meter Reset
2
Meter Hold
Peak Displ
Funct Reset
3
Meter Hold
Tare Action Tare Action
4
Peak Disp
Tare Action Funct Reset
5
Meter Reset Tare Action Meter Reset
6
Funct Reset Tare Action Meter Reset
7
Tare Reset
Tare Action Meter Reset
8
Displ Blank
Tare Action Meter Reset
9
Meter Reset Displ Blank Meter Reset
A
Funct Reset Displ Blank Meter Reset
B
Displ Item
Tare Action Tare Action
C
Displ Item
Displ Blank Funct Reset
D
Meter Reset Displ Item
Meter Reset
E
Funct Reset Displ Item
Meter Reset
F
Meter Hold
Displ Item
Meter Reset