10. ENABLING & LOCKING OUT MENU ITEMS

For security reasons and ease of meter operation, any and all
menu items may be disabled or "locked out" so that they are no
longer accessible from the front panel. Each function to be
disabled can be set to "1" in menu items Loc 1 through Loc 4, and
each function to be enabled can be set to "0." The Loc menu items
can in turn be locked out by installing an internal hardware jumper.
With the jumper installed, the operator only has access to enabled
menu items. With the jumper removed, the operator has access to
the Loc menu items.

SETTING THE HARDWARE LOCKOUT JUMPER

To access the lockout jumper, remove the rear panel per Section
10 and locate jumper “a” in the lower portion of the power supply
board next to the input connectors (see figure).

SETTING SOFTWARE LOCKOUTS

Set the lockout digit to "1" if you do not want the
menu item to be changed by an operator, other-
wise to “0”.

Loc 1 Loc 2 Loc 3 Loc 4

Press the

MENU key until Loc 1, Loc 2, Loc 3

or Loc 4 is displayed. Note: hardware lockout
jumper “a” must be removed (see above).

0

0000

12345
Press to set the flashing digit to "0" to enable the
menu item or to "1" to disable. Press MENU to
enter. See the table to the right for the list of menu
items that can be enabled or disabled.

1

1111

Press to display the lockout status, consisting of
1’s and 0’s. The left digit will flash. Press again
to step to the next digit, which will flash.

Lockout Digits for Menu Items

Loc 1

2 – Net/Gross display (reset key)
3 – Input type selection
4 – Setup, config. & decimal pt.
5 – Count-by and auto-zero

Loc 2

2 – Filtering
3 – Scale, Lo In, Hi In
4 – Offset, Lo Rd, Hi Rd
5 – Tare

Loc 3

2 – Alarm config.
3 – Change alarm setpoints
4 – Analog output scaling
5 – Serial comm. config.

Loc 4

2 – View peak value
3 – View alarm setpoints
4 – Front panel function resets
5 – Front panel meter reset