beautypg.com

20 future development – Impulse MAQ20 User Manual

Page 4

background image

Model Number

MAQ20-RTD31

MAQ20-RTD32

–200°C to +850°C (100Ω Pt), 3 selectable ranges

0°C to +850°C (100Ω Pt), 3 selectable ranges

–80°C to +300°C (120Ω Ni), 3 selectable ranges

0Ω to 5kΩ (Potentiometer), 3 selectable ranges

–200°C to +850°C (500Ω Pt), 3 selectable ranges

0°C to +850°C (1000Ω Pt), 3 selectable ranges

0°C to +160°C (10Ω Cu)

0°C to +160°C (50Ω Cu)

Number of Channels

Per Channel Setup

CMR

NMR

6

Individually configurable for range,

alarm limits, averaging

100dB at 50Hz or 60Hz

20dB at 50Hz or 60Hz

Accuracy

(1)

±0.06% Span

Bandwidth

Scan Rate

Alarms

Open Input Response

Power Supply Current

3Hz

200 Ch/s

High / High-High / Low / Low-Low

Upscale or Downscale

40mA

Specifications: 2- or 3-Wire RTD & Potentiometer

Input Modules

Typical at T

A

= +25°C and +24VDC system power

(1) Includes conformity, hysteresis and repeatability.

Model Number

MAQ20-BRDGN

MAQ20-BRDGW

Full bridge, Narrow bandwidth

Full bridge, Wide bandwidth

Number of Channels

Per Channel Setup

4

Individually configurable for range,

alarm limits, averaging

Input Range

Excitation

±10mV to ±100mV

10.0V

CMR

NMR (MAQ20-BRDGN)

100dB at 50Hz or 60Hz

20dB at 50Hz or 60Hz

Accuracy

(1)

±0.06% Span

Bandwidth

Scan Rate

Alarms

Power Supply Current

3Hz, 3kHz

200 Ch/s for 3Hz model

High / High-High / Low / Low-Low

400mA

Model Number

MAQ20-FREQ

500Hz to 100kHz

Number of Channels

Per Channel Setup

4

Individually configurable for range,

alarm limits

Zero Crossing Input

Min/Max Input

Hysteresis

Min Pulse Width

TTL Input

Min/Max Input

Hysteresis

Min Pulse Width

Excitation

100mVp-p/170Vp-p

±50mV

4µs

0.8V/2.4V

1.5V

4µs

+5V at 8mA

CMR

100dB at 50Hz or 60Hz

Accuracy

(1)

±0.05% Span

Scan Rate

Alarms

Power Supply Current

1500 Ch/s

High / High-High / Low / Low-Low

30mA

Specifications: Strain Gage Input Modules -

Preliminary

Typical at T

A

= +25°C and +24VDC system power

Specifications: Frequency Input Modules -

Preliminary

Typical at T

A

= +25°C and +24VDC system power

(1) Includes linearity/conformity, hysteresis and repeatability.

(1) Includes linearity/conformity, hysteresis and repeatability.

Controller Modules

CANbus

Wireless

Standalone
I/O Modules

True RMS Input

PID Loop Control

Single and Three Phase Monitoring

Ch-to-Ch Isolated Inputs

High Sample Rate / High Bandwidth Inputs

16 and 24 Bit Analog Input

DC and AC LVDT

Accelerometer Input

Two-Wire Transmitter Input

Serial Interface, RS-232 and RS-485

Interface to Existing Dataforth Signal Conditioning Modules

MAQ

®

20 Future Development

Flexible Backbone System Allows 4, 8, 16 and/or 24 Module Configuration in

19” Rack Space