Hach-Lange SC 60 User Manual
Page 25

21
Installation
3.5.1 Connecting the sc Sensor in a Non-hazardous Location
3.5.1.1 Attaching a sc Sensor with a Quick-connect Fitting
The sensor cable is supplied with a keyed quick-connect fitting for
easy attachment to the controller, see
Figure 17
. Retain the
connector cap to seal the connector opening in case the sensor
must be removed. Optional extension cables may be purchased to
extend the sensor cable length. If the total cable length exceeds
100 m (300 ft), a termination box must be installed. When used with
the termination box, the maximum cable length is 1000 m (3000 ft).
See the Replacement Parts section in the sensor manual for part
number information.
Figure 17 Attaching the Sensor using the Quick-connect Fitting
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAYC
RELAYB
RELAYA
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+
DATA
+
OUT2
– DATA
– OUT2
SERVICE REQUEST
SHIELD/CHASSIS GND
+
V
+
OUT1
GND
– OUT1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG
OUT
PU
TS
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAYC
RELAYB
RELAYA
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+
DATA
+
OUT2
– DATA
– OUT2
SERVICE REQUEST
SHIELD/CHASSIS GND
+
V
+
OUT1
GND
– OUT1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG
OUT
PU
TS
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
J5
J5