Installation 3.4.2 connecting the analog outputs, Disconnect power – Hach-Lange SC 60 User Manual
Page 23

19
Installation
3.4.2 Connecting the Analog Outputs
DANGER
Explosion hazard. Do not connect or disconnect electrical
components or circuits to the equipment unless power
has been switched off or the area is known to be
non-hazardous.
Two isolated analog outputs (1 and 2) are provided, see
Figure 16
on page 20
. Each output can be set to 0–20 or 4–20 mA, and can
be assigned to represent the measured parameter or secondary
measurement such as temperature. Make connections with
twisted-pair shielded wire and connect the shield at the controlled
component end or at the control loop end. Do not connect the
shield at both ends of the cable. Use of non-shielded cable may
result in radio frequency emission or susceptibility levels higher
than allowed. Maximum loop resistance is 500 Ohm. Refer to the
sensor manual for output software setup.