Operation, 1 menu structure – Hach-Lange GANIMEDE N User Manual
Page 37

37
Operation
Edit
Program
Create and save a new sample batch
Phosphorus
Definition of a sample batch for phosphorus
Name
Name, maximum 20 characters long, for the sample batch e.g. Well 1-10
Users
Name of the user. The usernames must be defined beforehand under the menu
option EDIT, USER.
Measurements
Single or double determination
Calibration
Two-point calibration with two selectable standard concentrations or
Multi-point calibration with access to standard curve 1, 2 or 3 (the standard curve
must first be recorded under the menu option ANALYSIS, CALIBRATION)
Sample data
Opens an input window for the description of individual samples.
Position
Automatic numbering (1–36)
Sample
Sample name, maximum 13 characters long
Method
Total phosphate or orthophosphate
Express as
PO
4
, PO
4
-P or P
2
O
5
Unit
Selection of the unit mg/L, g/L, µg/L, mmol/L, mg/100 g or Ext.
Control standard
A control standard with freely selectable concentration and permissible deviation can
also be measured on the sample tray.
Control standard selection YES or NO.
Dilution
The selection of a dilution factor against which the result is to be offset
Factor
Freely definable specific sample factor (0.001 – 9999)
Sampling
Date and time of sampling
Note
Note regarding the sample, maximum 30 characters long
Nitrogen
Definition of a sample batch for nitrogen
Name
Name, maximum 20 characters long, for the sample batch e.g. Well 1-10
User
Name of the user. The usernames must be defined beforehand under the menu
option EDIT, USER.
Measurements
Single or double determination
Calibration
Two-point calibration with two selectable standard concentrations in the low and high
measurement range respectively or
Multi-point calibration with access to standard curve 1, 2 or 3 (the standard curve
must first be recorded under the menu item ANALYSIS, CALIBRATION)
Sample data
Opens an input window for the description of individual samples.
Position
Automatic numbering (1–36)
Sample
Sample name, maximum 13 characters long
Method
Low measurement range (0.5–30 mg/L TN
B
) or high (30–150 mg/L TN
B
)
Express as
TN
Unit
Selection of the unit mg/L, g/L, µg/L, mmol/L, mg/100 g or Ext.
Control standard
A control standard with freely selectable concentration and permissible deviation can
also be measured on the sample tray.
Control standard selection YES or NO.
Dilution
The selection of a dilution factor against which the result is to be offset
Factor
Freely definable specific sample factor (0.001 – 9999)
Sampling
Date and time of sampling
Note
Note regarding the sample, maximum 30 characters long
4.2.1
Menu structure