2 installation of the tubes for the ganimede n, Installation – Hach-Lange GANIMEDE N User Manual
Page 23

23
Installation
Table 1 Tube overview for analysis unit P
3.1.4.2
Installation of the tubes for the Ganimede N
1. Install all tubes of the analysis unit in accordance with
.
2. Screw the large blue lid onto the empty waste bottle (1500 mL).
Note: The waste bottle can take the reagent and sample waste from 100 individual
determinations.
3. Connect the GCA 200 reagents with the corresponding lids of the same color.
4. Install sample tube A.
To do so, lead the free end of sample tube A out through the left-hand side of the
analysis unit housing.
Connect the upper end of the sample needle with sample tube A.
5. Insert the sample needle into the guide of the sample changer arm.
Note: The sample needle must not reach the bottom of the sample container when magnetic
rods are used.
Identification
Color in tube
plan
Length
Dia-
meter
Colored
adapter
Connected:
A: (included in
the scope of
delivery of the
sample
changer)
Black
400 mm
2.67 mm
-
Sample changer sample needle
-> Left-hand side of the analysis unit housing
-> Valve 1 connection f
B
Red
140 mm
2.67 mm
-
Valve 1 connection b
-> Left-hand analysis cuvette
C
Pink
380 mm
3.04 mm
Yellow/blue
Valve 1 connection g
-> Pump 1
-> Digestion cuvette
D
Gray
300 mm
3.04 mm
-
Digestion solution A
-> Valve 1 connection d
E
Green
530 mm
3.2 mm
Yellow/blue
Right-hand analysis cuvette
-> Pump 2
-> Valve 2 connection g
F
Blue
450 mm
3.2 mm
-
Color solution B
-> Valve 2 connection e
G
Green
300 mm
3.04 mm
-
Reduction solution C
-> Valve 2 connection c
H
Orange
180 mm
3.04 mm
-
Distilled water
-> Valve 2 connection b
J
Black
550 mm
3.53 mm
Gray/green/g
reen
Analysis cuvette
-> Pump 3
K
Purple
500 mm
4.17 mm
Gray/black/p
urple
Lower analysis cuvette
-> Pump 4
-> T-piece -\> Valve 1 connection c
and -\> Waste bottle
L
Black
300 mm
3.04 mm
-
Acid solution D
-> Valve 2 connection d