beautypg.com

Fluid Components International MT86_MT86HT Manual CE Conformance User Manual

Page 2

background image

FLUID COMPONENTS INTL

APPENDIX D - CE CONFORMANCE

Models MT86, MT86HT

D - 2

Doc. No. 003162 Rev. F

INSTALLATION CONFORMITY CRITERIA

Grounding

All enclosures must be grounded to earth ground through a path of less than 1 ohm.

Interconnecting Cables

All interconnecting cables between the instrument's local enclosure, remote enclosure, power source and mounting
device shall be enclosed in metal conduit. AC power input cabling shall be enclosed separately in conduit before
entrance into the instrument and is not to be combined with switch or monitor output cabling.

LOCATION OF CE MARK DOCUMENTATION

European Location

The technical documentation file part A resides at Fluid Components Intl, European Service Center, Beatrix
De Rijkweg 8, 5657 Eg Einhoven, Netherlands, Phone: 31-40-2-571-972 FAX: 31-40-2-517-809.

Manufacturer Location

The technical documentation file part B resides at the Configuration Management department of Fluid Components
Intl. 1755 La Costa Meadows Dr., San Marcos, CA 92069 USA, Phone: (800) 854-1993 FAX: (619) 736-6250.