Document 06EN003327 Rev. -
45
FlexCOR
™
Model CMF Series
6.1
Signal Transmitter
IP 67 and 19”
(terminal board
083H4260,
083H4253 &
083H4255)
Installation
1) Mains supply 115 to 230 V a.c. from building installation Class II. A switch or circuit-breaker
(max. 15 A) shall be included in the building installation. It must be in close proximity to the
equipment and within easy reach of the OPERATOR, and it shall be marked as the
disconnecting device for the equipment.
2) Protective conductor terminal. Required cable min. AGW16 or 1.5 Cu.
The insulation between the connected mains supply and 24 V a.c./d.c. supply for the flow-
meters, models 24 V a.c./d.c. shall at least be rated with double or reinforced insulation at
mains voltage.
For field wiring installation National Installation Code shall be met of the country, where the
flowmeters are installed.
Digital output
If the internal resistance of the loads exceeds 10K ohms, it is recommended to connect an
external 10K ohms load resistor in parallel to the load.
6.
Electrical connection