Fluid Components International CMF Series Manual Technical Data User Manual
Page 3
![background image](/manuals/330616/3/background.png)
Technical data
FlexCOR
ä
Model CMF Series
8
Fluid Components Intl
Document 06EN003327 Rev. -
Measurement of
Mass flow [lb/min / kg/s], volume flow [gpm, l/s], fraction [%],
°Brix, density [kg/m
3
], temperature [°F,
°C]
Current output
Current
0-20 mA or 4-20 mA
Load
< 800 ohm
Time constant
0-30 s adjustable
Digital output
Frequency
0-10 kHz, 50% duty cycle
Time constant
0-30 s adjustable
Active
24 V d.c., 30 mA,
1 KW
£ R
load
£ 10 KW, short-circuit-protected
Passive
3-30 V d.c., max. 110 mA,
1 KW
£ R
load
£ 10 KW
Relay
Type
Change-over relay
Load
42 V / 2 A peak
Functionality
Error level, error number, limit, direction
Digital input
11-30 V d.c., Ri = 13.6 K
W
Functionality
Start/hold/continue batch, zero point adjust, reset totalizer 1/2, force output, freeze output
Galvanic isolation
All inputs and outputs are galvanically isolated, isolation voltage 500 volts
Cut-off
Low-flow
0-9.9% of maximum flow
Limit function
Mass flow, volume flow, fraction, density, sensor temperature
Totalizer
Two eight-digit counters for forward, net or reverse flow
Display
Background illumination with alphanumerical text, 3 × 20 characters to indicate flow rate, totalized
values, settings and faults.
Operation: -4 to 122°F (
-20 to +50°C), max. rel. humidity 80% to 87°F (31°C) decreasing to 50% at 104°F
(40
°C) according to UL 3101.
During storage: -40 to 158°F (-40 to +70
°C) (Humidity max. 95%)
Communication
Prepared for client mounted add-on modules
Enclosure
Material
Standard 19" insert of aluminium/steel (DIN 41494)
Dimensions
Width:
21 TE
Height:
3 HE
Rating
IP 20 to IEC 529 and DIN 40050
Load
Version: 1 G, 1-800 Hz sinusoidal in all directions, to IEC 68-2-6
EMC performance
Emission
EN 50081-1 (Light industry)
Immunity
EN 50082-2 (Industry)
Namur
Within the value limits according to "Allgemeine Anforderung" with error criteria A in accordance with
NE 21
Supply voltage
24 V version
230 V version
Supply
24 V d.c./a.c., 50-60 Hz
115/230 V a.c., 50-60 Hz
Fluctation
24 V d.c.,
-25 to 25%
+10 to -10%
24 V a.c.,
-16 to 25%
Power consumption
10 W
26 VA
Fuse
230 V version: T400 mA, T 250V (IEC 127) - Not to be changed by user
24 V version: T1A, T 250V (IEC 127) - Not to be changed by user
Evironment
Environmental conditions acc. to UL 3101:
Indoor use
Altitude up to 6500 FT. (2000 m)
POLLUTION DEGREE 2
Ex approval
[EEx ia] IIC, DEMKO Ex 99E.125729X
Mass Flowmeter 19" IP 20
2.2.2 Mass Flowmeter 19" IP 20
The Transmitter is also available in the 19" version with outputs increased to 3 current outputs, 2 digital
outputs, 2 relay outputs, 1 digital input
Other data is identical to the above
Transmitter 19" insert
version with extended
outputs
2.2.3
Transmitter 19" IP 20 with extended outputs