Burkert Type 6624 User Manual
Page 2
1.
OPERATING INSTRUCTIONS
The operating instructions contain important information.
• Read the instructions carefully and follow the safety instructions
in particular.
• Keep the instructions in a location where they are available to
every user.
• The liability and warranty for Type 6624 are void if the operating
instructions are not followed.
2.
INTENDED USE
The rocker solenoid valve Type 6624 is designed for analytical,
medical and laboratory applications. It is predominantly used to dose,
fill, mix, and distribute liquids and gases.
• Do not use Type 6624 outside without taking suitable protective
measures.
• Do not use rectified AC voltage as the power supply without
smoothing.
• Operate only when in perfect condition and pay attention to correct
storage, transportation, installation and operation.
3.
BASIC SAFETY INSTRUCTIONS
These safety instructions do not make allowance for any contin-
gencies and events which may arise during installation, operation
and maintenance.
WARNING!
Danger – high pressure!
• Before loosening lines or valves, turn off the pressure and vent
the lines.
Risk of burns/risk of fire if used continuously through hot
device surface!
• Keep Type 6624 away from highly flammable substances and
media and do not touch with bare hands.
• Do not obstruct the heat release required for operation.
To prevent injury, ensure the following:
• Secure equipment/device from unintentional actuation.
• Do not use in potentially explosive area.
• Do not make any internal or external changes.
• Only trained technicians may perform installation and mainte-
nance work.
• After an interruption in the power supply, ensure that the pro-
cess is restarted in a controlled manner.
• Observe the general regulations of technology.
Electrostatic sensitive components/modules!
Observe the requirements in accordance with EN 61340-5-1 and
5-2 to minimize/avoid the possibility of damage caused by a sudden
electrostatic discharge!
4.
TECHNICAL DATA
4.1. Operating Conditions
Permitted temperatures for diaphragm material FFKM:
environment: +15 ... +55 °C, medium: +15 ... +50 °C
Extended temperature ranges: depending on the materials (see
data sheet).
Media: aggressive, neutral, gaseous and liquid media which do
not attack housing and seal materials (Recommendation: see
Check that resistance is adequate in individual cases.
Degree of protection: IP40
Protection class: 3 acc. to VDE 0580
4.2. Conformity
Type 6624 conforms with the EC Directives according to the EC
Declaration of Conformity.
4.3. Standards
The applied standards, which verify conformity with the EC Direc-
tives, can be found on the EC-Type Examination Certificate and /
or the EC Declaration of Conformity.
4.4. Electrical Data
See type label.
5.
INSTALLATION / DISASSEMBLY
5.1. Fluid Installation
WARNING!
Risk of injury from high pressure in the equipment.
• Before loosening lines or valves, turn off the pressure and vent
the lines.
Installation position: any position, preferably with actuator above.
→
Clean pipelines and flange connections.
→
Install dirt filter in the direction of flow in front of the valve (mesh
size 5 μm).
WARNING!
Risk of medium escaping if seal is seated incorrectly
• Make sure that the supplied seal fits properly.
• Use only connection plate of adequate quality and with a flat
surface.
Screws
Cover
Seals
Connection
plate
→
Drill holes according to
drilling pattern
(dimensions see data
sheet).
→
Remove cover (if fitted).
→
Insert the seals into the
valve.
→
Correctly assign fluid
connection configuration 1,
2 and 3 on valve and
connection plate.
→
Attach valve with screws
(Recommended tightening
torque 0,5 Nm ... 0,8 Nm).
→
Check installation for leaks.
5.2. Electrical Installation
Electrical power supply: Tolerance ± 10 %
Permissible residual ripple: Tolerance ± 10 % of the nominal
voltage
Power consumption: Start / switch-on pulse 4 W
Holding power
1 W
If the polarity is not correct, the valve will not function (see
data sheet).
5.3. Disassembly
WARNING!
Risk of injury from hazardous media.
• Before loosening lines or valves, flush out hazardous media,
depressurize and drain the lines.
6.
MAINTENANCE
Under normal operating conditions, the valve is maintenance-free.
7.
MALFUNCTIONS
If malfunctions occur, check
• that the fluid connections have been correctly assigned
according to the operating principles,
• whether the operating pressure is within the permitted range,
• the power supply and valve control
• the electric connections have the correct polarity.
8.
TRANSPORTATION, STORAGE, DISPOSAL
• Transport and store Type 6624 in shock-resistant packaging to
protect against moisture and dirt.
• Permitted storage temperature: -10 … +65 °C.
• Dispose of the device and packaging according to the appli-
cable disposal and environmental protection regulations.
Type 6624
english