CONFIGURE I/O MODULE 1

CFG MOD1 -- Configure I/O Module 1 (Choice)

YES

NO

This prompt allows access to the setup parameters associated with installation of the optional ISO-MOD
interface modules. If NO is selected, the unit will skip ahead to CFG MOD2. If YES is selected, configura-
tion and calibration of the module installed in the first position is accessible.

MODULE TYPE

MOD TYPE-- Module Type (Choice)

NONE

No Module Installed

4-20MA

4-20mA Analog Output

RATE

Rate Pulse Output

RELAY

Relay

Output

Select the type of module installed from the list.

ISO MOD 4-20 mA

ISO-MOD -- 4-20 mA

FLOW

4MA

FLOW

20MA

CAL

4MA

CAL

20MA

4-20

TEST

Configured via jumper selections for either a passive (current sinking) or active (current sourcing) trans-
mission mode (see Part 3 for details), the 4-20 mA Output Module interfaces with virtually all recording
and logging systems by transmitting an analog current signal that is proportional to system flow rate.
Independent 4 mA and 20 mA span settings are established in memory using the flow measuring range
entries. These entries can be set anywhere in the measuring range of the instrument. Output resolution
of the module is 12-bits (4096) discrete points) and the module can drive up to 800 Ohms of lead with its
internal 24 V isolated power source.

4-20 MA SPAN

The FLOW 4MA and FLOW 20MA entries are used to set the span of the 4-20 mA analog output. These
entries are volumetric rate units that are equal to the volumetric units configured as Engineering Rate
Units and Engineering Units Rate Interval.

For example, to span the 4-20 mA output from 0 GPM to +100 GPM, with 12 mA being 50 GPM, set the
FLOW 4MA and FLOW 20MA values as follows:

FLOW 4MA = 0.0

FLOW 20MA = 100.0