Select a transducer mounting location with adequate straight runs of pipe, both upstream and down-
stream, to achieve stable readings

1

. Examples of minimum upstream and downstream requirements are

included in Figure 2.2.

*

**

Flow

*

**

Flow

*

**

Flow

*

**

Flow

Flow

*

**

Flow

*

**

24

24

14

10

10

10

5

5

5

5

5

5

*

**

Upstream

Pipe

Diameters

Downstream

Pipe

Diameters

Piping Configuration

and Transducer Positioning

FIGURE 2.2 - UPSTREAM/DOWNSTREAM PIPE REQUIREMENTS

1

The DFX system will provide repeatable measurements on piping systems that do not meet these

requirements, but the accuracy may be influenced to various degrees.