B&K Precision 889B - Manual User Manual
Page 7

6
2. AC Voltage Measurement (True RMS):
Range
: 2V, 20V, 200V, and 600V
Resolution
: 1mV, 10mV, 100mV, and 1V
Accuracy
: +/- (0.8% + 5 digits)
Input Impedance
: 1 M-Ohm
3. DC Current Measurement:
Range
: 2mA, 20mA, 200mA, and 2000mA
Resolution
: 1uA, 10uA, 100uA, and 1mA
Accuracy
: +/- (0.4% + 3 digits)
Current Shunt
: 0.1 Ohm @ >20mA, 10 Ohm @
≤20mA
4. AC Current Measurement (True RMS):
Range
: 2mA, 20mA, 200mA, and 2000mA
Resolution
: 1uA, 10uA, 100uA, and 1mA
Accuracy
: +/- (0.8% + 5 digits)
Current Shunt
: 0.1 Ohm @ >20mA, 10 Ohm @
≤20mA
Note:
The accuracy of DC/AC voltage/current measurements is only applied when in 5% - 100% of the range.
5. LCR Measurement:
Z Accuracy (Ae):
|Zx|
Freq.
20M ~
10M
(
Ω)
10M ~
1M
(
Ω)
1M ~
100K
(
Ω)
100K ~
10K
(
Ω)
10K ~
1K
(
Ω)
1K ~ 100
(
Ω)
100 ~ 1
(
Ω)
1 ~ 0.1
(
Ω)
DCR
2% ±1
1% ±1 0.5% ±1 0.2% ±1 0.1% ±1 0.2% ±1 0.5% ±1 1% ±1
100Hz
120Hz
1KHz
10KHz
5% ±1
2% ±1
100KHz
200KHz
NA
5% ±1
2% ±1
1% ±1 0.4% ±1 1% ±1
2% ±1
5% ±1
Note:
1. The accuracy applies when the test level is set to 1Vrms.
2. Ae multiplies 1.25 when the test level is set to 250mVrms.
3. Ae multiplies 1.50 when the test level is set to 50mVrms.
4. When measuring L and C, multiply Ae by
2
1 Dx
+
if the Dx>0.1.
: Ae is applied only when the test level is set to 1Vrms.