4 dc resistance measurement, 5 ac impedance measurement, 6 capacitance measurement – B&K Precision 889B - Manual User Manual
Page 17: 7 inductance measurement, Dc resistance measurement, Ac impedance measurement, Capacitance measurement, Inductance measurement

16
2.2.4
DC Resistance Measurement
The DC resistance measurement measures the resistance of an unknown component by 1VDC. Press the
L/C/Z/DCR key to select the DCR measurement. The LCD will display:
2.2.5
AC Impedance Measurement
The AC impedance measurement measures the Z of an unknown device. Press the L/C/Z/DCR key to select the
Z measurement. The LCD will display:
The testing level and frequency can be selected by pressing the Level key and Freq key, respectively.
2.2.6
Capacitance Measurement
To measure the capacitance of a component, users may be able to press the L/C/Z/DCR key to select either Cs
(Serial Mode) or Cp (Parallel Mode) measurement mode. If the serial mode (Cs) is selected, the D, Q and ESR
can be shown on the secondary display. If the parallel mode (Cp) is selected, only the D and Q can be shown on
the secondary display. The following shows some examples of capacitance measurement:
The testing level and frequency can be selected by pressing the Level key and Freq key, respectively.
2.2.7
Inductance Measurement
Press the L/C/Z/DCR key to select Ls or Lp mode for measuring the inductance in serial mode or parallel mode.
If the serial mode (Ls) is selected, the D, Q and ESR can be shown on the secondary display. If the parallel mode
(Lp) is selected, only the D and Q can be shown on the secondary display. The following shows some examples
of inductance measurement:
The testing level and frequency can be selected by pressing the Level key and Freq key, respectively.