beautypg.com

Main specifications optional accessories – Atec Yokogawa-DL9000 Series User Manual

Page 5

background image

This active probe can be used in
combination with the DL9000 series to
measure signals with an analog
bandwidth up to 1.5 GHz.

Bandwidth:

DC to 2.5 GHz (-3 dB)

Attenuation and DC accuracy:

10:1 (

±2%)

Input resistance:

100 k

Ω (±2%)

Input capacitance: Approx. 0.9 pF (typ.)
Dynamic range:

±7 V

PBA2500 2.5 GHz active probe

Input resistance:

10 M

Ω ±2%

(when used with the DL9000)

Input capacitance: Approx. 14 pF (typ.)

(when used with the DL9000)

Attenuation:

Fixed to 1/10

Bandwidth:

DC to 500 MHz
(within -3 dB)

Max. Input voltage:

±600 V DC + AC peak

This DC block is used to remove bias
voltage occurring when the PBL5000
probe is used.

Overall length:

Approx. 25 mm

Connector type:

SMA

Input impedance: 50

Frequency range: 20 MHz to 6 GHz
Max. input voltage:

±10 V (DC + ACpeak)

PB500 500 MHz passive probe

This wideband low capacitance probe
can be used with the 50 ohm input
setting.

Connector type:

SMA

Input resistance:

450

Ω or 950 Ω

Input capacitance: Approx. 0.25 pF

(typ. 450

Ω),

0.4 pF (typ. 950

Ω)

Attenuation:

10:1 or 20:1

Bandwidth:

DC to 5 GHz (-3 dB)

Max. input voltage: 20 Vrms, 40 V ACpeak

PBL5000 5 GHz low capacitance probe

701975 50 ohm DC block

Bandwidth:

DC to 500 MHz
(within -3 dB)

Attenuation:

1/10 (fixed)

Input impedance (typ.):

100 k

Ω/2.5 pF

Max. allowable differential voltage:

±12 V (DC + ACpeak)

Max. common mode input voltage:

±30 V (DC + ACpeak)
(Output impedance: 50

Ω)

701920

±12 V/500 MHz differential probe

Bandwidth:

DC to 100 MHz (-3 dB)

Max. continuous input range:

30 Arms

Amplitude accuracy:

0 to 30 Arms:

±1% of rdg ±1 mV

Up to 50 Apeak:

±2.0% of rdg (DC, 45 to 66 Hz)

Weight:

Approx. 240 g

701932 DC to 100 MHz 30 Arms current probe

Bandwidth:

DC to 100 MHz (-3 dB)

Attenuation:

1/10 or 1/100 (selectable)

Max. allowable differential voltage:

±700 V (DC + ACpeak)

Max. common mode input voltage:

±700 V (DC + ACpeak)
(common to both 1/10 and
1/100 attenuation ratios)

701921

±700 V/100 MHz differential probe

This differential probe is suited for
observation of fast differential signals,
such as LVDS. Using this probe in
combination with the DL9000 series,
you can observe differential signals with
an analog bandwidth up to 1.5 GHz.

Bandwidth:

DC to 2.0 GHz (-3 dB)

Attenuation and DC accuracy:

10:1 (50

Ω)

Input capacitance: Approx. 1.1 pF (typ.)
Max. differential input voltage:

±5 V

PBD2000 2.0 GHz differential probe

Bandwidth:

DC to 200 MHz (-3 dB)

Attenuation:

1/10 (fixed)

Max. allowable differential voltage:

±20 V (DC + ACpeak)

Max. common mode input voltage:

±60 V (DC + ACpeak)

Output impedance: 50

701922

±20 V/200 MHz differential probe

Bandwidth:

DC to 2 MHz (-3 dB)

Max. continuous input range:

500 Arms

Amplitude accuracy:

0 to 500 Arms:

±1% of rdg ±5 mV

Up to 700 Apeak:

±2.0% of rdg (DC, 45 to 66 Hz)

Weight:

Approx. 520 g

701931 DC to 20 MHz 500 Arms current probe

9

8

Main Specifications

Optional Accessories

Models

Model name (No.)

Max. sampling rate

Freq. BW

Max. record length

DL9040 (701307)

5 GS/s

500 MHz

2.5 MW

DL9040L (701308)

5 GS/s

500 MHz

6.25 MW

DL9140 (701310)

5 GS/s

1 GHz

2.5 MW

DL9140L (701311)

5 GS/s

1 GHz

6.25 MW

DL9240 (701312)

10 GS/s

1.5 GHz

2.5 MW

DL9240L (701313)

10 GS/s

1.5 GHz

6.25 MW

Basic Specifications

Input channels:

4 (CH1 to CH4)

Input coupling:

AC, DC, GND, DC50

Input impedance:

1 M

Ω ±1.0% approx. 20 pF (when using PB500 probe, 10

M

Ω ±2.0%, approx. 14 pF)

50

Ω ±1.5%

Voltage axis sensitivity:

For 1 M

Ω input : 2 mV/div to 5 V/div (steps of 1-2-5) ranges

For 50

Ω input : 2 mV/div to 500 mV/div (steps of 1-2-5)

Maximum input voltage:

For 1 M

Ω input : 150 Vrms CAT I

For 50

Ω input : 5 Vrms or less and 10 Vpeak or less

DC offset max. setting range:

For 1 M

Ω input

(When probe attenuation set to 1:1)

2 mV/div to 50 mV/div

:

±1 V

100 mV/div to 500 mV/div :

±10 V

1 V/div to 5 V/div

:

±100 V

For 50

Ω input

2 mV/div to 50 mV/div

:

±1 V

100 mV/div to 500 mV/div :

±5 V

Vertical (voltage) axis sensitivity:

DC accuracy

1

:

For 1 M

Ω input : ±(1.5% of 8 div + offset voltage accuracy)

For 50

Ω input : ±(1.5% of 8 div + offset voltage accuracy)

Offset voltage axis accuracy

1

: 2 mV/div to 50 mV/div

:

±(1% of setting + 0.2 mV)

100 mV/div to 500 mV/div :

±(1% of setting + 2 mV)

1 V/div to 5 V/div

:

±(1% of setting + 20 mV)

Voltage standing-wave ratio (VSWR): 1.5 or less within frequency bandwidth (typical value

4

)

Frequency characteristics

1, 2

(Attenuation point of -3 dB when inputting a sinewave of amplitude

±2 div or equivalent)

For 50

Ω input

DL9040/9040L

DL9140/DL9140L

DL9240/DL9240L

0.5 V/div to 10 mV/div:

DC to 500 MHz

DC to 1 GHz

DC to 1.5 GHz

5 mV/div:

DC to 400 MHz

DC to 750 MHz

DC to 1 GHz

2 mV/div:

DC to 400 MHz

DC to 600 MHz

DC to 750 MHz

For 1 M

Ω input (from the probe tip when using the PB500 dedicated passive probe)

5 V/div to 10 mV/div:

DC to 500 MHz

DC to 500 MHz

DC to 500 MHz

5 mV/div to 2 mV/div:

DC to 400 MHz

DC to 400 MHz

DC to 400 MHz

Residual noise level

3

:

0.4 mV rms or 0.05 div rms, whichever is larger (typical value

4

)

A/D conversion resolution:

8-bit (25 LSB/div)

Bandwidth limit:

For each channel, select from FULL, 200 MHz, 20 MHz, 8
MHz, 4 MHz, 2 MHz, 1 MHz, 500 kHz, 250 kHz, 125 kHz,
62.5 kHz, 32 kHz, 16 kHz, and 8 kHz (separately configurable
on each of channels CH1 to CH4); Limit implemented with
analog (200 MHz, 20 MHz) and digital filters (IIR+ FIR).

Max. sampling rate:

DL9040/9040L/9140/9140L

DL9240/9240L

Real time sampling mode:

Interleave mode ON:

5 GS/s

10 GS/s

Interleave mode OFF:

2.5 GS/s

5 GS/s

Repetitive sampling mode:

2.5 TS/s

2.5 TS/s

Maximum record length

DL9040/9140/9240

DL9040L/DL9140L/DL9240L

2.5 MW

6.25 MW

Time axis setting range:

500 ps/div to 50 s/div (steps of 1-2-5)

Time base accuracy

1

:

±0.001%

Time axis measurement accuracy

1

:

± (0.01% + 10 ps + 1 sample interval)

Max. acquisition rate

5

:

When using 1.25 MW, 60 waveforms/sec/ch
When using 12.5 kW, 9000 waveforms/sec/ch
When using 2.5 kW, 25000 waveforms/sec/ch

Min. dead time (N single)

5

:

400 ns or less (equivalent to 2.5 M waveforms/sec)

Trigger Section

Trigger modes:

Auto, Auto Level, Normal, Single, and N Single

Trigger source:

CH1 to CH4:

Signals applied to measurement input terminals

LINE:

Connected commercial power signal (only available with
Edge trigger)

EXT:

Signal input from EXT TRIG IN terminal

Trigger level range:

CH1 to CH4:

±4 divisions from the screen center

EXT:

±2 V (1:1), ±20 V (10:1 when used with a probe)

Trigger level setting resolution:

CH1 to CH4:

0.01 div

EXT:

5 mV (1:1), 50 mV (10:1 when used with a probe)

Window comparator:

Separately configurable on each of channels CH1 to CH4

Center:

±4 divisions from the screen center

Width:

±4 divisions from Center

Trigger level accuracy

CH1 to CH41:

±(0.2 div + 10% of trigger level)

EXT

1

:

±(50 mV + 10% of trigger level)

Trigger sensitivity:

DL9040/DL9040L

DL9140/DL9140L

DL9240/DL9240L

CH1 to CH41

1 divp-p

DC to 500 MHz

DC to 1 GHz

DC to 1 GHz

EXT

1

100 mVp-p

DC to 100 MHz

DC to 100 MHz

DC to 100 MHz

where Edge OR1 1 divp-p

DC to 50 MHz

DC to 50 MHz

DC to 50 MHz

Trigger types:

Edge/State

Edge:

Trigger occurs on the edge of a single trigger source.

Edge (Qualified):

Trigger occurs on the edge of a single trigger source when
Qualification condition is true.

Edge OR:

Trigger occurs on the OR logic of the edge conditions set
to multiple trigger sources.

State:

Trigger occurs on ENTER/EXIT when the state condition
is true.

Width

Pulse:

Trigger occurs on a width of a single trigger source.

Pulse (Qualified):

Trigger occurs on a width of a single trigger source when
Qualification condition is true.

Pulse State:

Trigger occurs on a width when the state condition is true.

Time width setting mode:

More than:

Trigger occurs upon change in condition when the condition
remains true longer than time T1.

Less than:

Trigger occurs upon change in condition when the condition
remains true shorter than time T1.

Between:

Trigger occurs upon change in condition when the condition
remains true longer than time T1 and shorter than time
T2.

Out of Range:

Trigger occurs upon change in condition when the condition
remains true shorter than time T1 and longer than time
T2.

Time out:

Trigger occurs when the condition is true for duration longer
than time T1.

Specified time (T1/T2):

1 ns to 10 s, 500 ps resolution

Time accuracy:

±(0.2% of setting + 1 ns)

Event Interval

Event Cycle:

Trigger occurs when the event cycle is within the specified
time range.

Event Delay:

After Event 1 occurs, trigger occurs on 1st occurrence of
Event 2 that satisfies the timing constraints. The trigger
process is reset if Event 1 or Event 2 occurs before the
timing constraints are satisfied.

Event Sequence:

After Event 1 occurs, trigger occurs on 1st occurrence of
Event 2 that satisfies the timing constraints. The trigger
process is reset if Event 1 occurs before the timing
constraints are satisfied.

Time width setting mode:

Function identical to the time width setting mode for Width

Specified time (T1/T2):

1.5 ns to 10 s, 500 ps resolution

Time accuracy:

±(0.2% of setting + 1 ns)

Event types:

Events can be selected from Edge, Edge Qualified, State,
Pulse, Pulse Qualified, Pulse State, I

2

C, CAN, SPI, and

Serial trigger types.

Enhanced:

TV: Trigger occurs on video signals of various broadcasting system formats

Mode:

NTSC, PAL, HDTV, USER

Input CH:

CH1-CH4

Sync Guard:

Hsync 60 to 90% (increments of 1%)

Line:

5-1054 (NTSC), 2-1251 (PAL), 2-1251 (HDTV), 2-2048
(USER)

Field:

1/2/X

Frame Skip:

1/2/4/8

I

2

C:

Triggers on I2C bus signals

Mode:

NON ACK, Every Start, General Call, (Start byte/HS
Mode), ADR&DATA

SPI:

Triggers on SPI (serial peripheral interface) bus signals

Mode:

3 wire, 4 wire

CAN:

Bit rate:

1 Mbps, 500 kbps, 250 kbps, 125 kbps, 83.3 kbps
User (freely settable in 100 bps increments)

Input channel:

CH1 to CH4: Input through differential probe

Mode:

SOF, Frame ID, Data field, Remote Frame, Error Frame,
Ack etc.

Serial pattern:

Triggers on general-purpose serial communication signals.

Max. bit rate:

50 Mbps

Max. bit length:

128 bits

Display

Display:

8.4-inch (21.3 cm) color TFT liquid crystal display

Display screen size:

170.5 mm (width)

ϫ 127.9 mm (height)

Total number of pixels:

1024

ϫ 768 (XGA)

Waveform display resolution:

800

ϫ 640