Tr i g g e r S y s t e m
TDS7054
TDS7104
TDS7154/TDS7254
TDS7404
Sensitivity
Internal DC Coupled
0.35 div DC to 50 MHz
0.35 div DC to 50 MHz
0.35 div DC to 50 MHz
0.35 div DC to 50 MHz
increasing to 1 div at 500 MHz
increasing to 1 div at 1 GHz
increasing to 1.5 div at 3 GHz
increasing to 1.5 div at 3 GHz
External (Auxiliary Input)
400 mV from DC to 50 MHz
250 mV from DC to 50 MHz
250 mV from DC to 50 MHz
250 mV from DC to 50 MHz
increasing to 750 mV
increasing to 500 mV
increasing to 350 mV
increasing to 350 mV
at 100 MHz
at 100 MHz
at 500 MHz
at 500 MHz
Main Trigger Modes
Auto, Normal and Single
Trigger Sequences
Main, Delayed by Time, Delayed by Events. All sequences can include separate horizontal delay
after the trigger event to position the acquisition window in time
Trigger Characteristics
Standard Trigger Types
Edge, Glitch, Runt, Width, Transition time, Timeout, Pattern, State, Setup/Hold
Communications-related Triggers
Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ
(requires Option SM)
encoded communications signals. Select between isolated positive or
negative one, zero pulse form or eye patterns as applicable to standard
Serial Pattern Trigger
32 bit serial word recognizer, bits specified in binary
(requires Option ST)
(high, low, don’t care) or hex format. Trigger on NRZ-encoded data up to 1.25 GBaud
Trigger Level Range
Internal
±12 divisions from center of screen
External (Auxiliary In)
±8 V
Line
fixed at 0 V
Trigger Coupling
DC, AC (attenuates <60 Hz), HF Rej (attenuates >30 kHz), LF Rej
(attenuates <80 kHz), Noise Reject (reduce sensitivity)
Trigger Holdoff Range
250 ns minimum to 12 s maximum