beautypg.com

Atec Tektronix-TDS7154 User Manual

Page 7

background image

Digital Phosphor Oscilloscopes

TDS7000 Series

Digital Phosphor Oscilloscopes •

www.tektronix.com/tds7000

7

Tr i g g e r S y s t e m

TDS7054

TDS7104

TDS7154/TDS7254

TDS7404

Sensitivity

Internal DC Coupled

0.35 div DC to 50 MHz

0.35 div DC to 50 MHz

0.35 div DC to 50 MHz

0.35 div DC to 50 MHz

increasing to 1 div at 500 MHz

increasing to 1 div at 1 GHz

increasing to 1.5 div at 3 GHz

increasing to 1.5 div at 3 GHz

External (Auxiliary Input)

400 mV from DC to 50 MHz

250 mV from DC to 50 MHz

250 mV from DC to 50 MHz

250 mV from DC to 50 MHz

increasing to 750 mV

increasing to 500 mV

increasing to 350 mV

increasing to 350 mV

at 100 MHz

at 100 MHz

at 500 MHz

at 500 MHz

Main Trigger Modes

Auto, Normal and Single

Trigger Sequences

Main, Delayed by Time, Delayed by Events. All sequences can include separate horizontal delay

after the trigger event to position the acquisition window in time

Trigger Characteristics

Standard Trigger Types

Edge, Glitch, Runt, Width, Transition time, Timeout, Pattern, State, Setup/Hold

Communications-related Triggers

Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ

(requires Option SM)

encoded communications signals. Select between isolated positive or

negative one, zero pulse form or eye patterns as applicable to standard

Serial Pattern Trigger

32 bit serial word recognizer, bits specified in binary

(requires Option ST)

(high, low, don’t care) or hex format. Trigger on NRZ-encoded data up to 1.25 GBaud

Trigger Level Range

Internal

±12 divisions from center of screen

External (Auxiliary In)

±8 V

Line

fixed at 0 V

Trigger Coupling

DC, AC (attenuates <60 Hz), HF Rej (attenuates >30 kHz), LF Rej

(attenuates <80 kHz), Noise Reject (reduce sensitivity)

Trigger Holdoff Range

250 ns minimum to 12 s maximum