Digital storage oscilloscope – Atec Tektronix-TDS6804B User Manual
Page 7
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
7
Digital Storage Oscilloscope
TDS6000B/C Series
Maximum Time Duration Captured at Highest Real-time Resolution (All Channels)
TDS6000B Models
TDS6000C Models
(All Channels)
(Two Channels)
Max Duration with Standard Memory
100 µs
100 µs
Max Duration with Opt. 2M
400 µs
400 µs
Max Duration with Opt. 3M
800 µs
800 µs
Max Duration with Opt. 4M
1.6 ms
1.6 ms
Acquisition Modes
All Models
Waveform Database
Accumulate Waveform Database providing three-dimensional
array of amplitude, time and counts
Sample
Acquire sampled values
Peak Detect
Captures narrow glitches at all real-time sampling rates
Minimum Peak Detect Pulse Width
50 ps
Averaging
From 2 to 10,000 waveforms included in average
Envelope
From 2 to 2x10
9
waveforms included in min-max envelope
Hi-Res
Real-time boxcar averaging reduces random noise and increases resolution
FastFrame
™
Acquisition
Acquisition memory divided into segments; maximum trigger rate >310,000
waveforms per second. Time of arrival recorded with each event
Roll Mode
200 kS/s at 8 Mb Record length and 500 kS/s at
4 Mb Record Length on all channels
Pinpoint
™
Trigger System
All Models
Sensitivity
Internal DC Coupled
0.5 div DC to 50 MHz increasing to 2.5 div at 7 GHz (TDS6000B models,
typical) 0.4 div DC to 50 MHz, increasing to 1 div at 3 GHz,
increasing to 3 div at 9 GHz (TDS6000C models, typical)
External (auxiliary input)
250 mV DC to 50 MHz, increasing to 350 mV at
1 GHz, increasing to 500 mV at 2 GHz (typical)
Trigger Characteristics
Main Trigger Modes
Auto, Normal and Single
A Event and Delayed B Event Trigger Types
Edge, Glitch, Runt, Width, Transition Time, Timeout, Pattern, State,
Setup/Hold, Window – all except Edge, Pattern and State
can be Logic State qualified by up to two channels
Trigger Sequences
Main, Delayed by Time, Delayed by Events, Reset by Time, Reset by State,
Reset by Transition. All sequences can include separate horizontal delay
after the trigger event to position the acquisition window in time
Communications-related Triggers
Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ encoded communications
(requires Option SM)
signals up to 3.125 Gb/s. Select among isolated positive or negative one,
zero pulse form or eye patterns as applicable to standard