Model d111-1, Model d301x, Model d5kv – Atec Thermo-ECAT-LTS User Manual
Page 8: Model dpi-1
Model D111-1
5 KA Current Probe suitable for all waveforms except WF5B
General Specifications
Sensitivity
0.1 Volt/Ampere +1/-0%
Output resistance
50 Ohms
Scope Coupling
1 MΩ DC/AC
Maximum peak current
5,000 Amperes
Useable rise time
25 nanoseconds
Low frequency 3dB point
5 Hz
High frequency 3dB point
15 MHz
Operating temperature
0°C to +65°C
Physical Specifications
Output connector
BNC
Weight
1.7 kg (3.7 lb)
Model D301X
50KA Current Probe suitable for WF5A, 5B
General Specifications
Sensitivity
0.01 Volt/Ampere +1/-0%
Output resistance
50 Ohms
Scope Coupling
1 MΩ DC/AC
Maximum peak current
50,000 Amperes
Useable rise time
200 nanoseconds
Low frequency 3dB cut-off
5 Hz
High frequency 3dB cut-off
2 MHz
Operating temperature
0°C to +65°C
Physical Specifications
Output connector
UHF (SO-239)
Weight
7.9 kg (17.5 lb)
Model D5KV
Voltage Probe suitable for all Waveforms to 5KV
General Specifications
Max. input voltage
5 kV
Scope Coupling
1 MΩ DC/AC
System attenuation
100:1, +/-2%
System input resistance
50 MΩ
Input capacitance
< 6 pF
System BW (-3 dB)
400 MHz
Operating temperature
0 °C to +50 °C
Physical Specifications
Cable length
2 m
Output connector
BNC
Model DPI-1
Probe accessory kit suitable for all Waveforms to 5KV
General Specifications
Safety Socket Dolphin Clips
Safety Socket Crocodile Clips
Safety Socket Pointed Probes
Safety Socket Grabber Probes
Safety Socket Lug Terminals
Safety Socket Clip-On
Safety Socket Cables
Safety Socket Leads
Safety Socket Plugs
T
her
m
o
Sc
ient
if
ic
EC
AT
Li
ght
ni
n
g
T
es
t
Sy
s
tem
(
LT
S)