Atec Gigatronics-8650A Series User Manual
Page 7

MODULATED POWER
MEASUREMENTS
The Giga-tronics 80400A Series
Modulated Power Sensors let you
measure the average power of ampli-
tude modulated, burst modulated and
other complex modulated signals — such
as TDMA signals — at bandwidths up to
40 kHz.
The Giga-tronics 80600A Series Modulated
Power Sensors provide bandwidth up to
1.5 MHz to measure the peak and average
power of CDMA signals.
Sensor Model
Signal Type
80301A
80350A
80401A
80601A
80701A
CW Power Level
–70 to +20 dBm
–30 to +20 dBm
–67 to +20 dBm
–67 to +20 dBm
–64 to +20 dBm
Amplitude Modulation
N/A
N/A
f
m
< 40 kHz, –60 to +20 dBm
f
m
< 1.5 MHz, –60 to +20 dBm
f
m
< 10 MHz, –60 to +20 dBm
Rate, Power Range
f
m
> 40 kHz, –60 to –20 dBm
f
m
> 1.5 MHz, –60 to –20 dBm
Two-Tone N/A
N/A
< 40 kHz, –60 to +20 dBm
< 1.5 MHz, –60 to +20 dBm
≤
10 MHz, –60 to +20 dBm
Maximum Separation
> 40 kHz, –60 to –20 dBm
> 1.5 MHz, –60 to –20 dBm
> 10 MHz, –60 to –20 dBm
Between Carriers
Pulse Modulation
N/A
> 350 ns
> 200 µs
> 300 µs
> 100 µs
Pulse Width
Pulse Width
Pulse Width
Pulse Width
Burst with Modulation
N/A
N/A
f
m
< 40 kHz, > 200 µs
f
m
< 1.5 MHz, > 300 µs
f
m
< 10 MHz, > 100 µs
f
m = modulation rate
Pulse Width; –40 to +20 dBm
Pulse Width; –40 to +20 dBm
Pulse Width; –30 to +20 dBm
f
m
> 40 kHz, > 200 µs
f
m
> 1.5 MHz, > 300 µs
f
m
> 10 MHz, > 100 µs
Pulse Width; –40 to –20 dBm
Pulse Width; –40 to –20 dBm
Pulse Width; –30 to –20 dBm
Sensor Measurement Capabilities
And the Giga-tronics 80701A Modulated
Power Sensor operating with the 8650A
power meter, provides system bandwidth
up to 10 MHz to measure the peak and
average power of wide band, third-genera-
tion CDMA signals over an 80 dB range.