UPC CONTROLLER SPECIFICATIONS

The UPC controller is essentially a 3f AC arbitrary waveform generator and Precision AC metering system. Each
waveform stored in the UPC is encoded with 12 bit amplitude and 10 bit time resolution for each cycle. The waveform for
each phase may be independently selected and may be independently
varied in amplitude and phase angle with respect to phase A.
The UPC output metering samples the output volts and amps at 512 samples per measurement using a 12 bit A/D converter.
This technique provides exceptional metering accuracy and resolution (20 bits), and delivers a high-fidelity waveform back
to a host computer for analysis.
The UPC includes a remote GPIB interface compatible with IEEE 488.2 and SCPI. An available option is an RS-232 serial
port that operates up to 38.4 kBaud.

Frequency:

20.00 to 5000 Hz

±

0.01%

Voltage:

Programmable, 0-VMAX, in 0.1 volt steps

Direct

(see table on page 6)

Voltage:

Multi-range units are equipped with

Transformer output transformers. When alternate range is

selected, voltage at transformer output is
programmable in steps of 0.5 volts.

Accuracy:

Executive voltage is within

±

50 mv (0.03%)

Command

of command voltage, referenced to the

Voltage

internal voltmeter with CSC engaged.

Accuracy:

±

0.01%, 20-5000 Hz

Output Zo:

Dynamic output impedance (Zo) is

(Optional)

programmable, 0 to

±

Zo max. in 0.1%

steps. Zo value in milliohms varies with
different models but usually results in a

±

10% change in output voltage at maximum
load amps.

Phase:

Phase Angle (f) of Phases B and C

Angle

relative to Phase A is programmable from 0-
359

°

in 1

°

increments

±

0.5

°

.

Current:

Current limit is programmable from 0

Limit

to Ipeak maximum of the power source.
Accuracy is

±

1%, resolution

±

0.05%.

Library:

Stores up to 99 steady state parameter

Steady

sets consisting of waveform, voltage,

State

frequency, f angle and current limit. Can be

Programs

executed by program number from the front
panel or the bus.

Library:

Stores up to 99 transient programs - one

Transient

associated with each steady state program.

Programs

Allows for changes in volts and frequency

vs. time, or waveform changes by cycle count.

Library:

Stores up to 99 waveforms that can be edited

Waveform

and executed in any manner and in any
output phase.

Voltmeter:

Range -

0-354 volts L-N
0-708 volts L-L

Resolution -

0.10VAC to front panel
0.001VAC to remote interface

Accuracy -

±

0.25% of reading

±

0.1% of

range (50-500Hz)

Ammeter:

Range -

300% of system current rating

Resolution -

0.01AAC to front panel
0.001AAC to remote interface

Accuracy -

±

0.25% of reading

±

0.1% of

range (50-500Hz)

Power Meter: Range -

Based on ammeter range

Resolution -

1.0 watts or VA to front panel
0.001 Watts or VA to remote
interface

Accuracy -

±

1% of Full Scale

Power Factor: Calculated and displayed to three significant
& Crest Factor digits.

Ext. Input:

Each phase is algebraically summed with
UPC waveform and amplified 25X to the
direct output.

Amplitude:

±

10 volt input for each phase

Mod. Input

modulates the output

±

100%

Sync Outputs:1) Zero crossing, Phase A

2) Transient start-stop
3) True when Transient is enabled
4) Clock - 1024 times the output freq.

Command:

Average time to start of parameter change

Response

from bus command (end of string character)

Time

is 50 ms. Ramp transition time to final value
is settable from 250

µ

s to 300 sec.

Waveform :

Permits waveform creation by entering %

Synthesis

amplitude and phase angle for the 2nd
through the 51st harmonics.

Waveform:

Reports voltage and current waveform

Analysis

harmonic content in % and phase angle
for the 2nd through the 51st harmonics.
Displays THD, OHD, EHD in %.

7