Metering waveform control/analysis, Function key provides access to special functions, Special functions accessed through upc setup menu – Atec Pacific-Power-AMX Series User Manual
Page 4
METERING
WAVEFORM CONTROL/ANALYSIS
V/I METER: ENTRY: 120.0
FREQ=60.00
Va=120.0
Vb=120.0
Vc=120.0
SENSE=INT
Vab=208.0 Vbc=208.0 Vca=208.0
MANUAL MODE Ia=06.00
Ib=06.22
Ic=06.15
POWER METER: PHASE A PHASE B PHASE C
KVA
0.720
0.746
0.738
KW
0.720
0.746
0.738
PF
1.000
1.000
1.000
AMPS METER: PHASE A PHASE B PHASE C
RMS
0.720
0.746
0.738
PEAK
1.044
1.119
1.383
CREST FACTOR
1.45
1.50
1.90
EDIT WAVEFORM: NUMBER=16
RANGE=2-22
STARTING PHASE ANGLE=0
0-359.5
°
ENDING PHASE ANGLE=0
0-359.5
°
VOLTAGE IN PERCENT=-100
(+/-)0-100%
WAVEFORM SYNTHESIS: WAVEFORM #2
HARMONIC:
2nd
3rd
4th
5th
6th
CONTENT:
.1%
0%
0%
0%
0%
ØANGLE:
0
°
0
°
0
°
0
°
0
°
ØA CURRENT THD=17.8 % OHD=17.8EHD=0.3%
HARMONIC:
2nd
3rd
4th
5th
6th
CONTENT:
.1%
17.8%
0%
0%
0%
ØANGLE:
0
°
0
°
0
°
0
°
0
°
FUNCTION KEY
PROVIDES ACCESS TO SPECIAL FUNCTIONS
SETUP: PRESS
1 FOR PROGRAM SETUP
2 FOR WAVEFORM SETUP
3 FOR GENERAL SETUP
4 FOR CALIBRATION MENU
PROGRAM SETUP
• Copy a program
• Delete a program
• Erase all memory,
reset CPU
WAVEFORM SETUP
• Edit a waveform
• Copy a waveform
• Waveform synthesis
GENERAL SETUP
• UPC setup
• LCD setup
• UPC status
• Power source status
• Range control
• Slew rate setup
CALIBRATION MENU
• Execute externally
referenced calibration
• View calibration
constants
SPECIAL FUNCTIONS ACCESSED
THROUGH UPC SETUP MENU
• SENSE
Establishes either local or external
sense for metering and CSC.
• CSC
Continuous Self Calibration –
provides for exceptional voltage
accuracy.
• PROGRAM
Programmable output impedance
Z
°
dynamically compensates for
output transformer or line
distribution losses. Can simulate a
soft power grid.
• TRANSITION Permits control of the transition
TIME
time when changing the output
voltage and frequency.
• FREQUENCY Sets min and max programmable
LIMITS
frequency limits.
• VOLTAGE
Sets min and max programmable
LIMITS
voltage limits.