5 Fluke Corporation Fluke 430 Series II Three-Phase Energy and Power Quality Analyzers

Product specifications

Model

Measurement range

Resolution

Accuracy

Volt
Vrms (ac+dc)

434-II

1 V to 1000 V phase to neutral

0.1 V

± 0.5 % of nominal voltage****

435-II and 437-II

1 V to 1000 V phase to neutral

0.01 V

± 0.1 % of nominal voltage****

Vpk

1 Vpk to 1400 Vpk

1 V

5 % of nominal voltage

Voltage Crest Factor (CF)

1.0 > 2.8

0.01

± 5 %

Vrms½

434-II

1 V to 1000 V phase to neutral

0.1 V

± 1 % of nominal voltage

434-II and 435-II

0.1 V

± 0.2 % of nominal voltage

Vfund

434-II

1 V to 1000 V phase to neutral

0.1 V

± 0.5 % of nominal voltage

435-II and 437-II

0.1 V

± 0.1 % of nominal voltage

Amps (accuracy excluding clamp accuracy)
Amps (ac +dc)

i430-Flex 1x

5 A to 6000 A

1 A

± 0.5 % ± 5 counts

i430-Flex 10x

0.5 A to 600 A

0.1 A

± 0.5 % ± 5 counts

1mV/A 1x

5 A to 2000 A

1A

± 0.5 % ± 5 counts

1mV/A 10x

0.5 A A to 200 A (ac only)

0.1 A

± 0.5 % ± 5 counts

Apk

i430-Flex

8400 Apk

1 Arms

± 5 %

1mV/A

5500 Apk

1 Arms

± 5 %

A Crest Factor (CF)

1 to 10

0.01

± 5 %

Amps½

i430-Flex 1x

5 A to 6000 A

1 A

± 1 % ± 10 counts

i430-Flex 10x

0.5 A to 600 A

0.1 A

± 1 % ± 10 counts

1mV/A 1x

5 A to 2000 A

1A

± 1 % ± 10 counts

1mV/A 10x

0.5 A A to 200 A (ac only)

0.1 A

± 1 % ± 10 counts

Afund

i430-Flex 1x

5 A to 6000 A

1 A

± 0.5 % ± 5 counts

i430-Flex 10x

0.5 A to 600 A

0.1 A

± 0.5 % ± 5 counts

1mV/A 1x

5 A to 2000 A

1A

± 0.5 % ± 5 counts

1mV/A 10x

0.5 A A to 200 A (ac only)

0.1 A

± 0.5 % ± 5 counts

Hz
Hz

Fluke 434 @ 50 Hz nominal

42.50 Hz to 57.50 Hz

0.01 Hz

± 0.01 Hz

Fluke 434 @ 60 Hz nominal

51.00 Hz to 69.00 Hz

0.01 Hz

± 0.01 Hz

Fluke 435/7 @ 50 Hz nominal

42.500 Hz to 57.500 Hz

0.001 Hz

± 0.01 Hz

Fluke 435/7 @ 60 Hz nominal

51.000 Hz to 69.000 Hz

0.001 Hz

± 0.01 Hz

Fluke 437 @ 400 Hz nominal

340.0 Hz to 460.0 Hz

0.1 Hz

± 0.1 Hz

Power
Watts (VA, var)

i430-Flex

max 6000 MW

0.1 W to 1 MW

± 1 % ± 10 counts

1 mV/A

max 2000 MW

0.1 W to 1 MW

± 1 % ± 10 counts

Power factor (Cos j/DPF)

0 to 1

0.001

± 0.1 % @ nominal load conditions

Energy
kWh (kVAh, kvarh)

i430-Flex 10x

Depends on clamp scaling and V nominal

± 1 % ± 10 counts

Energy loss

i430-Flex 10x

Depends on clamp scaling and V nominal

± 1 % ± 10 counts

Excluding line resistance accuracy

Harmonics
Harmonic order (n)

DC, 1 to 50 Grouping: Harmonic groups according to IEC 61000-4-7

Inter-harmonic order (n)

OFF, 1 to 50 Grouping: Harmonic and Interharmonic subgroups according to IEC 61000-4-7

Volts

%f

0.0 % to 100 %

0.1 %

± 0.1 % ± n x 0.1 %

%r

0.0 % to 100 %

0.1 %

± 0.1 % ± n x 0.4 %

Absolute

0.0 to 1000 V

0.1 V

± 5 % *

THD

0.0 % to 100 %

0.1 %

± 2.5 %

Amps

%f

0.0 % to 100 %

0.1 %

± 0.1 % ± n x 0.1 %

%r

0.0 % to 100 %

0.1 %

± 0.1 % ± n x 0.4 %

Absolute

0.0 to 600 A

0.1 A

± 5 % ± 5 counts

THD

0.0 % to 100 %

0.1 %

± 2.5 %

Watts

%f or %r

0.0 % to 100 %

0.1 %

± n x 2 %

Absolute

Depends on clamp scaling and V nominal

± 5 % ± n x 2 % ± 10 counts

THD

0.0 % to 100 %

0.1 %

± 5 %

Phase Angle

-360° to +0°

± n x 1°