Coupling impedance
Surge voltage parameters under open circuit conditions
18μF
9μF+10Ω
Front time
1.2μs±30%
1.2μs±30%
Time to half value
Current rating < 25 A
50μs+10μs/-10μs
50μs+10μs/-25μs
Current rating 25 A – 60 A
50μs+10μs/-15μs
50μs+10μs/-30μs
Current rating 60 A – 100 A
50μs+10μs/-20μs
50μs+10μs/-35μs
NOTE The measurement of the surge voltage parameters should be done with the power supply input port of the coupling/decoupling network open-circuit.
Voltage waveform specification at the EUT port of the coupling/decoupling network
Coupling impedance
Surge current parameters under short circuit conditions
18μF
9μF+10Ω
Front time
8μs±20%
2.5μs±30%
Time to half value
20μs±20%
25μs±30%
NOTE The measurement of the surge current parameters should be done with the power supply input port of the coupling/decoupling network open-circuit.
Current waveform specification at the EUT port of the coupling/decoupling network
4. Test Setup
■
Injection to power supply lines
Wooden table
EUT
SURGE OUT
LINE OUT
Isolation transformer
AC
PE
Surge generator
Injecting 1.2/50 combination wave put out from CDN of
LSS-F02 to battery drive of a PC.
Floating circuit is adapted according to the Standard for
the output.
In LSS-F02 series, enable the automated operation with
the program function along the above figure.
■
Injection to telecom lines and power supply lines
Injecting 1.2/50 combination wave and telecom surge
(10/700 combination wave) put out from CDN of
LSS-F02 to power supply lines and telecom lines of a
fax machine.
In LSS-F02 series, enable the automated operation with
the program function along the above figure.
SURGE OUT
LINE OUT
Isolation transformer
Wooden table
AC
PE
Surge generator
TEL LINE OUT
EUT
* The above setup figure is one example of the test setup our lightning surge simulator.
There is no specification on the Standard.