Option
Maximum capacity is single phase AC240V / 30A
and the dielectric strength is 4kV
Available for AC lines power supply in LSS-F02
series and widely usable for the other various noise
generators or measurement equipments.
Isolation Transformer
MODEL:TF-2302P
Parameter
Specification
Max. input voltage Single phase 240V MAX (50 / 60Hz)
Max. output voltage 30A MAX
Dielectric strength Primary to core : AC4kV (1 minute)
Secondary to core : AC4kV (1 minute)
Primary – Secondary : AC4kV (1 minute)
Insulation resistance ≧100MΩ at DC500V
Dimension / Mass
(W)350×(H)475×(D)400mm
(Eye bolts and handles excluded)
Approx. 60 kg
Parameter
Specification
Max. input voltage Single phase / 3-hase 240V MAX (50 / 60Hz)
Transformer connection Y-connection
Max. output voltage 50A MAX
Dielectric strength Primary to core : AC4kV (1 minute)
Secondary to core : AC4kV (1 minute)
Primary – Secondary : AC4kV (1 minute)
Insulation resistance ≧100MΩ at DC500V
Dimension / Mass
(W)500×(H)640×(D)700mm
(Eye bolts and handles excluded)
Approx. 350 kg
Maximum capacity is 3-phase AC600V / 50A and
the dielectric strength is 4kV
Available for AC lines power supply in LSS-F02
series and widely usable for the other various noise
generators or measurement
Isolation Transformer
MODEL:TF-6503P
Warning Lamp
MODEL
:
11-00008A
Usable together with LSS-F02 series. The blinking
makes the operators or neighbors pay attention to the
test processing.
Protective Safety Fence
MODEL
:
11-00010A
Enable to materialize the safe test environment
with connection to interlock function equipped in
LSS-F02 series. The safety measure can be sure
together with the EUT protective safety box.
Relay terminal board to connect output of LSS-F02
series to EUT. Enable to connect any outlet figure in
the world when wiring to the attached multi-outlet.
For single phase 3 lines (Voltage resistible capacity
4.5kV)
* The same option for 3-phase 5 lines also available
(18-00058B)
Terminal Connection Board attached with Multi-Outlet (3p)
MODEL:18-00048B
EUT Protective Safety Box
MODEL
:
11-00005A/11-00006A
Protection box to prevent access to EUT during the
test.
Further safety is secured together with the safety
protective fence
Optical conversion adaptor Used for remote control
with PC. 5m of optical fiber cable with USB interface
attached.
Optical USB module
MODEL:07-00022A
Used for the surge test to interconnection lines defined in
IEC61000-4-5 Standard. The EUT power capacity is DC50V/1A
and enable to inject the surge to interconnection lines up to
6,600V. Possible to bypass inductor (20mH) with connecting
the attached connection plug to inductor bypass terminal in
DC output. Possible to equip the attached surge protective
arrestor between each line and ground.
* The conversion (05-H1784)cable is needed additionally.
CDN for Interconnection Lines
MODEL:LSS-INJ6400SIG
Parameter
Specification
Surge input voltage
500V~6.600V (Combination wave)
EUT power capacity
DC50V/1A
Max. line number
4 lines
Decoupling coil
20mH each line
Impedance matching resistor 40Ω ±10%
Dimension / Mass
(W)488×(H)456×(D)550mm Approx. 45kg
Waveform Pre-Checking Cables Set
MODEL:05-00099A
Fixtures for checking voltage waveforms and current
waveforms of LSS-F02 series.
Followings are necessary for the checking additionally.
・Oscilloscope (Differential operation function built-in)
・High voltage probes (for surge voltage measurement /
Voltage resistibility necessary)
・Current probe (For surge short current measurement)
・Isolation transformer (for oscilloscope)
・Earth cable (for PE connection)
ch1
ch2
Surge generator
High voltage probe X 2 pcs.
Connect GND of high voltage probe
Grounding
conductor
Isolation
transformer
PE noncontact
Power supply
HOTch1ch
2PECOMSurge
Oscilloscope which
enables differential
measurement
* Both capacitor coupling and arrestor coupling are possible in LSS-INJ6400SIG. Diode coupling which is defined in IEC61000-4-5 ed.2 (2005) newly
is a test method which assumes the injection of dozens of voltage level surge to interconnection lines. Please contact us if low voltage surge injection is
required since the surge injection in LSS-INJ6400SIG can be more than 90V.