Specification

Parameter

Specification

Note

Surge generating unit

1.2/50μs-8/20μs

Output voltage

0.5kV ~ 15kV ±10%

Combination waveforms

Front time

1.2μs ±30%

Common for the all models

Duration

50μs ±20%

Voltage step : 0.1kV step

Output current

250A ~ 7500A ±10%

The setting can be from 0kV

Front time

8μs ±20%

Duration

20μs ±20%

10/700μs-5/320μs

Output voltage

0.5kV ~ 15kV ±10%

Combination waveforms

Front time

10μs ±30%

Models:C1/C3

Duration

700μs ±20%

Voltage step : 0.1kV step

Output current

12.5A ~ 375A ±10%

The setting can be from 0kV

Front time

5μs ±20%

Duration

320μs ±20%

Output polarity

Positive / Negative

Interval

10 sec.~989 sec., depending on the set voltage 10 sec. (< 6kV)

15 sec.~ in 10/700μs waveform

Output impedance

2Ω ±10%

1.2/50μs waveform

40Ω ±10%

10/700μs waveform

AC/DC CDN

Coupling surge waveform

1.2/50μs-8/20μs combination waveforms

Max. coupling surge voltage / current

Up to the values which can be set

Coupling network

18μF

Between LINE - LINE (10Ω +9μF selectable)

Correspondent to IEC61000-4-5

10Ω ±9μF

Between LINE - PE (18μF selectable)

Injection mode

Between LINE - LINE, Between LINE - PE

Power supply lines structure for EUT

Single phase AC

L/N/PE

Model : A1 / C1

DC

+/-/PE

3-phase AC

L1/L2/L3/N/PE(Common for single phase and 3-phase)

Model : A3 / C3

DC

+/-/PE

EUT power capacity

AC240V/20A MAX 50/60Hz、DC125V/20A MAX

Model : A1 / C1

AC500V/50A MAX 50/60Hz、DC125V/50A MAX

Model : A3 / C3

Decoupling coil

1.5mH

Phase angle control

0 ~ 360° ±10°

CDN for Telecom lines (Only in model C1 and C3)

Coupling surge waveform

1.2/50μs-8/20μs combination waveforms

10/700μs-5/320μs combination waveforms

Max. coupling surge voltage / current

Up to the values which can be set

Impedance matching resistors

40Ω

80Ωper 1 line at 2 lines

1.2/50 μs waveform

160Ωper 1 line at 4 lines

25Ω

0Ωper 1 line at 2 lines

10/700 μs waveform

100Ωper 1 line at 4 lines

Coupling mode

Common mode / Normal mode

Coupling network

Gas arrestor : 90V

Line for EUT

2 lines / 4 lines DC50V/100mA MAX

Selectable

Decoupling coil

20mH

Others

Voltage monitor

BNC output, 1/2000±10%

In open-circuit for SURGE OUT

Current monitor

BNC output, 1mV/A±10%

In short-circuit for SURGE OUT

External communication

RS-232C optical communication

Power supply

AC100V ~ AC240V ±10% 50Hz / 60Hz

Dimension

W555×H1800×D790 mm

Projection excluded (in all models)

Mass

Approx. 300kg

Model : A1 / A3

Approx. 340kg

Model : C1 / C3

Item

Specification / Function

Q'ty

Correspondent model

Surge output cable

HOT / COM

2 pcs.

Common

Output cable to power supply lines For single phase : L / N / PE

3 pcs.

A1/C1

For 3-phase : L1 / L2 / L3 / N / PE

5 pcs.

A3/C3

Output cable to telecom lines

For 1~4 lines and GND

5 pcs.

C1/C3

Arrestor unit

For coupling : Equipped to main unit panel

4 pcs.

C1/C3

For input protection : Equipped to main unit panel

4 pcs.

Cable for monitor

BNC-BNC cable

1 pc.

Common

External interlock connector

5P plug (Short between #1 - #3)

1 pc.

Common

Power supply cable

For AC100V, 3P equipped with G connector cable

1 pc.

Common

High voltage connector cap

Equipped to main unit panel

5 pcs.

A1/C1

7 pcs.

A3/C3

FG cable

For grounding the body

1 pc.

Common

Instruction manual

1 volume

Common

Standard accessory

These products use parts containing mercury. Please comply with lows or regulation in countries or states the products are used for the disposal.

Certain periodical inspection shall be recommended since consumable parts are contained in the products.

 

In the test to 3-phase 5 lines (with PE) power supply lines, a message which alert the inspection per around 200 sets (in the test to single phase (with PE) power supply lines, it is done per around 800 sets).

 

(1 set in this case means that the test shall be done with 2 levels (eg. 0.5kV and 1kV) for the test series according to IEC61000-4-5)

* Exchange timing of the parts may differed depending on the operative conditions and environment. Please contact us for the more details.

LSS-F02-□□

How to understand the model numbers

A : 1.2/50μs-8/20μs (Totally 1 kind surge generates)

C : 1.2/50μs-8/20μs、10/700μs-5/320μs (Totally 2 kinds)

1 : Model for single phase EUT L/N/PE

3 : Model for 3-phase EUT L1/L2/L3/N/PE(Available both for single phase & 3-phase)

LSS-F02 series specification