Lss-f02 series specification – Atec Noiseken-LSS-F02 Series User Manual
Page 5
w w w . n o i s e k e n . c o m
5
Specification
Parameter
Specification
Note
Surge generating unit
1.2/50μs-8/20μs
Output voltage
0.5kV ~ 15kV ±10%
Combination waveforms
Front time
1.2μs ±30%
Common for the all models
Duration
50μs ±20%
Voltage step : 0.1kV step
Output current
250A ~ 7500A ±10%
The setting can be from 0kV
Front time
8μs ±20%
Duration
20μs ±20%
10/700μs-5/320μs
Output voltage
0.5kV ~ 15kV ±10%
Combination waveforms
Front time
10μs ±30%
Models:C1/C3
Duration
700μs ±20%
Voltage step : 0.1kV step
Output current
12.5A ~ 375A ±10%
The setting can be from 0kV
Front time
5μs ±20%
Duration
320μs ±20%
Output polarity
Positive / Negative
Interval
10 sec.~989 sec., depending on the set voltage 10 sec. (< 6kV)
15 sec.~ in 10/700μs waveform
Output impedance
2Ω ±10%
1.2/50μs waveform
40Ω ±10%
10/700μs waveform
AC/DC CDN
Coupling surge waveform
1.2/50μs-8/20μs combination waveforms
Max. coupling surge voltage / current
Up to the values which can be set
Coupling network
18μF
Between LINE - LINE (10Ω +9μF selectable)
Correspondent to IEC61000-4-5
10Ω ±9μF
Between LINE - PE (18μF selectable)
Injection mode
Between LINE - LINE, Between LINE - PE
Power supply lines structure for EUT
Single phase AC
:
L/N/PE
Model : A1 / C1
DC
:
+/-/PE
3-phase AC
:
L1/L2/L3/N/PE(Common for single phase and 3-phase)
Model : A3 / C3
DC
:
+/-/PE
EUT power capacity
AC240V/20A MAX 50/60Hz、DC125V/20A MAX
Model : A1 / C1
AC500V/50A MAX 50/60Hz、DC125V/50A MAX
Model : A3 / C3
Decoupling coil
1.5mH
Phase angle control
0 ~ 360° ±10°
CDN for Telecom lines (Only in model C1 and C3)
Coupling surge waveform
1.2/50μs-8/20μs combination waveforms
10/700μs-5/320μs combination waveforms
Max. coupling surge voltage / current
Up to the values which can be set
Impedance matching resistors
40Ω
80Ωper 1 line at 2 lines
1.2/50 μs waveform
160Ωper 1 line at 4 lines
25Ω
5
0Ωper 1 line at 2 lines
10/700 μs waveform
100Ωper 1 line at 4 lines
Coupling mode
Common mode / Normal mode
Coupling network
Gas arrestor : 90V
Line for EUT
2 lines / 4 lines DC50V/100mA MAX
Selectable
Decoupling coil
20mH
Others
Voltage monitor
BNC output, 1/2000±10%
In open-circuit for SURGE OUT
Current monitor
BNC output, 1mV/A±10%
In short-circuit for SURGE OUT
External communication
RS-232C optical communication
Power supply
AC100V ~ AC240V ±10% 50Hz / 60Hz
Dimension
W555×H1800×D790 mm
Projection excluded (in all models)
Mass
Approx. 300kg
Model : A1 / A3
Approx. 340kg
Model : C1 / C3
Item
Specification / Function
Q'ty
Correspondent model
Surge output cable
HOT / COM
2 pcs.
Common
Output cable to power supply lines For single phase : L / N / PE
3 pcs.
A1/C1
For 3-phase : L1 / L2 / L3 / N / PE
5 pcs.
A3/C3
Output cable to telecom lines
For 1~4 lines and GND
5 pcs.
C1/C3
Arrestor unit
For coupling : Equipped to main unit panel
4 pcs.
C1/C3
For input protection : Equipped to main unit panel
4 pcs.
Cable for monitor
BNC-BNC cable
1 pc.
Common
External interlock connector
5P plug (Short between #1 - #3)
1 pc.
Common
Power supply cable
For AC100V, 3P equipped with G connector cable
1 pc.
Common
High voltage connector cap
Equipped to main unit panel
5 pcs.
A1/C1
7 pcs.
A3/C3
FG cable
For grounding the body
1 pc.
Common
Instruction manual
-
1 volume
Common
Standard accessory
●
These products use parts containing mercury. Please comply with lows or regulation in countries or states the products are used for the disposal.
●
Certain periodical inspection shall be recommended since consumable parts are contained in the products.
In the test to 3-phase 5 lines (with PE) power supply lines, a message which alert the inspection per around 200 sets (in the test to single phase (with PE) power supply lines, it is done per around 800 sets).
(1 set in this case means that the test shall be done with 2 levels (eg. 0.5kV and 1kV) for the test series according to IEC61000-4-5)
* Exchange timing of the parts may differed depending on the operative conditions and environment. Please contact us for the more details.
LSS-F02-□□
How to understand the model numbers
A : 1.2/50μs-8/20μs (Totally 1 kind surge generates)
C : 1.2/50μs-8/20μs、10/700μs-5/320μs (Totally 2 kinds)
1 : Model for single phase EUT L/N/PE
3 : Model for 3-phase EUT L1/L2/L3/N/PE(Available both for single phase & 3-phase)
●
LSS-F02 series specification
●