Level – Atec Rohde-Schwarz-ESL Series User Manual
Page 5

Version 02.00, March 2013
Rohde & Schwarz R&S
®
ESL EMI Test Receiver
5
Level
Display range
displayed noise floor to +20 dBm
Maximum rated input level
DC voltage
50 V
CW RF power
30 dBm (= 1 W)
Peak RF power
36 dBm (= 4 W) <3 s
Max. pulse voltage
150 V
Max. pulse energy
pulse width 10 µs
10 mWs
Intermodulation
Third-order intermodulation
intermodulation-free dynamic range, level 2 × –20 dBm, reference level –10 dBm
f
in
< 30 MHz
>54 dBc (TOI +7 dBm, typ. +12 dBm)
f
in
≥ 30 MHz
>60 dBc (TOI +10 dBm, typ +18 dBm)
Second harmonic intercept (SHI)
20 MHz ≤ f
in
≤ 3 GHz
nominal +35 dBm
1 dB compression of input mixer
0 dB RF attenuation, preamplifier = OFF,
f > 200 MHz
nominal +5 dBm
Displayed average noise level (DANL, analyzer mode)
0 dB RF attenuation, termination 50 Ω, RBW = 1 kHz, VBW = 1 Hz,
sample detector, log scaling, tracking generator OFF, normalized to 1 Hz
frequency
preamplifier = OFF
9 kHz to 1 MHz
<–115 dBm (1 Hz)
1 MHz to 10 MHz
<–120 dBm (1 Hz)
10 MHz to 50 MHz
<–130 dBm (1 Hz)
50 MHz to 3 GHz
<–140 dBm (1 Hz)
3 GHz to 5 GHz
<–136 dBm (1 Hz)
5 GHz to 6 GHz
<–130 dBm (1 Hz)
with R&S
®
FSL-B22 option
frequency
preamplifier = ON
9 kHz to 1 MHz
<–130 dBm (1 Hz)
1 MHz to 10 MHz
<–135 dBm (1 Hz)
10 MHz to 50 MHz
<–145 dBm (1 Hz)
50 MHz to 3 GHz
<–152 dBm (1 Hz)
3 GHz to 5 GHz
<–146 dBm (1 Hz)
5 GHz to 6 GHz
<–140 dBm (1 Hz)
frequency
preamplifier = ON, typical values
500 MHz
–162 dBm (1 Hz)
1 GHz
–160 dBm (1 Hz)
3 GHz
–158 dBm (1 Hz)
6 GHz
–147 dBm (1 Hz)