Measurement specifications – Atec Fluke-1586A User Manual
Page 5
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 5
1586-2586 High-Capacity Module specifications
1586-2588 DAQ-STAQ Multiplexer specifications
General
Maximum input
50 V
Offset voltage
<2 μV
3-wire internal resistance mismatch
<50 mΩ
Basic CJC accuracy
0.25 °C
General
Maximum input
50 V
Offset voltage
<2 μV
3-wire internal resistance mismatch
<50 mΩ
Basic CJC accuracy
0.6 °C
LAN
Function
Control and data transfer
Network protocols
Ethernet 10/100, TCP/IP
Command protocol
SCPI
RS-232
Connector
D-sub 9 pin (DE-9)
Baud rates
1200, 2400, 4800, 9600, 19200, 38400
Function
Temperature source control output
Dimensions
Height
150 mm
Width
245 mm
Depth
385 mm
Weight
6 kg (typical configuration)
Shipping weight
9.5 kg (typical configuration)
Conformity
CE, CSA, IEC 61010 3rd edition
Accuracy specifications generally apply with medium and slow sample rates (unless otherwise noted), after a warm-up
time of 1 hour, and within an environment temperature range of 18 °C to 28 °C, and may depend on the channel.
The confidence level for accuracy specifications is 95 % within one year of calibration.
Measurement specifications
Scan rate
Fast
10 channels per second max (0.1 seconds per channel)
Medium
1 channel per second (1 second per channel)
Slow
4 seconds per channel
Display resolution
4 ½ to 6 ½ digits, depending on function and Sample Rate (see Measurement
Characteristics tables below to find the display resolution of temperature readings)
PRT/RTD
Temperature range
−200 °C to 1200 °C (depending on the sensor)
Resistance range
0 Ω to 4 kΩ
Offset compensation
0 Ω to 400 Ω, 4-wire
automatic current reversal
400 Ω to 4000 Ω or 3-wire
none
Source current reversal interval
(0 Ω to 400 Ω range)
Fast sample rate
2 ms
Medium sample rate
250 ms
Slow sample rate
250 ms
Maximum lead resistance (4-wire Ω)
2.5 % of range per lead for 400 Ω and 4 kΩ ranges.