10 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner

Thermocouple measurement characteristics

Range

Temperature Display Resolution

Slow / Medium Sample Rate

Fast Sample Rate

−270 °C to 2315 °C

0.01 °C

0.1 °C

DC voltage

Maximum input

50 V on any range

Common mode rejection

140 dB at 50 Hz or 60 Hz (1 kΩ unbalance in LOW lead) ± 50 V peak maximum

Normal mode rejection

55 dB for power line frequency ± 0.1 %, ± 120 % of range peak maximum

A/D linearity

2 ppm of measurement + 1 ppm of range

Input bias current

30 pA at 25 °C

DC voltage accuracy

Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for Channel 1, medium or slow sample

rate. For channels x01 through x20 or when using Fast sample rate, add the numbers given in the table to the accuracy

specification. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers

by the temperature deviation and add to the accuracy specification.

Range

Accuracy

Channel 1

Ch. x01 – x20

Fast

Sample Rate

T.C./ °C Outside

18 °C to 28 °C

±100 mV

0.0037 % + 0.0035 %

add 2 μV

add 0.0008 % of range

0.0005 % + 0.0005 %

±1 V

0.0025 % + 0.0007 %

add 2 μV

add 0.0008 % of range

0.0005 % + 0.0001 %

±10 V

0.0024 % + 0.0005 %

add 0.0008 % of range

0.0005 % + 0.0001 %

±50 V

0.0038 % + 0.0012 %

add 0.0008 % of range

0.0005 % + 0.0001 %

DC voltage input characteristics

Range

Resolution

Input Impedance

Slow / Medium

Fast

±100 mV

0.1 μV

1 μV

10 GΩ

1

±1 V

1 μV

10 μV

10 GΩ

1

±10 V

10 μV

100 μV

10 GΩ

1

±50 V

100 μV

1 mV

10 MΩ ±1 %

1

Input beyond ±12 V is clamped. The clamp current is up to 3 mA.

DC current accuracy

Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for medium or slow sample rate.

When using a fast sample rate, add the number given in the table to the accuracy specification. If the environment temper-

ature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to

the accuracy specification.

DC current

Input protection

0.15 A resettable PTC