Thermocouple measurement characteristics
Range
Temperature Display Resolution
Slow / Medium Sample Rate
Fast Sample Rate
−270 °C to 2315 °C
0.01 °C
0.1 °C
DC voltage
Maximum input
50 V on any range
Common mode rejection
140 dB at 50 Hz or 60 Hz (1 kΩ unbalance in LOW lead) ± 50 V peak maximum
Normal mode rejection
55 dB for power line frequency ± 0.1 %, ± 120 % of range peak maximum
A/D linearity
2 ppm of measurement + 1 ppm of range
Input bias current
30 pA at 25 °C
DC voltage accuracy
Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for Channel 1, medium or slow sample
rate. For channels x01 through x20 or when using Fast sample rate, add the numbers given in the table to the accuracy
specification. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers
by the temperature deviation and add to the accuracy specification.
Range
Accuracy
Channel 1
Ch. x01 – x20
Fast
Sample Rate
T.C./ °C Outside
18 °C to 28 °C
±100 mV
0.0037 % + 0.0035 %
add 2 μV
add 0.0008 % of range
0.0005 % + 0.0005 %
±1 V
0.0025 % + 0.0007 %
add 2 μV
add 0.0008 % of range
0.0005 % + 0.0001 %
±10 V
0.0024 % + 0.0005 %
–
add 0.0008 % of range
0.0005 % + 0.0001 %
±50 V
0.0038 % + 0.0012 %
–
add 0.0008 % of range
0.0005 % + 0.0001 %
DC voltage input characteristics
Range
Resolution
Input Impedance
Slow / Medium
Fast
±100 mV
0.1 μV
1 μV
10 GΩ
1
±1 V
1 μV
10 μV
10 GΩ
1
±10 V
10 μV
100 μV
10 GΩ
1
±50 V
100 μV
1 mV
10 MΩ ±1 %
1
Input beyond ±12 V is clamped. The clamp current is up to 3 mA.
DC current accuracy
Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for medium or slow sample rate.
When using a fast sample rate, add the number given in the table to the accuracy specification. If the environment temper-
ature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to
the accuracy specification.
DC current
Input protection
0.15 A resettable PTC