Electric and Magnetic Field Measurement

efa-300 field analyzer

Table 1: Spectrum FFT Sensitivity (Noise / Spurious)

magnetiC (b-) fielD

eleCtriC (e-) fielD

100 cm

2

Probe

internal Probe

3 cm Probe

2 kHz Range

<45 nT @

≤48 Hz

<4 nT @ >48 Hz

<0.05 nT @ noise floor

<400 nT @

≤48 Hz

<42 nT @ >48 Hz

<2 nT @ noise floor

<260 nT @

≤48 Hz

<23 nT @ >48 Hz

<0.2 nT @ noise floor

<0.3 V/m @

≤48 Hz

<0.1 V/m @ >48 Hz

<0.02 V/m @ noise floor

32 kHz Range

<2 nT @ <200 Hz

<0.3 nT @ 200 Hz to 20 kHz

<0.6 nT @ >20 kHz

<0.07 nT @ noise floor

<22 nT @ <200 Hz

<11 nT @ 200 Hz to 20 kHz

<11 nT @ >20 kHz

<1.5 nT @ noise floor

<10 nT @ <200 Hz

<2 nT @ 200 Hz to 20 kHz

<3 nT @ >20 kHz

<0.3 nT @ noise floor

<0.1 V/m @

≤20 kHz

<3 V/m @ >20 kHz

<0.05V/m @ noise floor

b-fielD unit

e-fielD moDule

Display

LCD Dot Matrix 128x64 Pixel with Backlight

Via B-Field Unit

Alarm, Adjustable Threshold

Acoustical, Optical

Via B-Field Unit

Current Documentation

(Specific Modes Only)

Input of Prevailing and Reference Current Value;

Storage with Measurement Value of Field

N/A

Interface (Remote Control, Data Memory)

Optical, Serial (RS-232)

Operating Temperature Range

0°C to +50°C

Humidity

<95% or <29 g/m³ Occasional Brief Condensation Tolerable

Operating

Interval, typical

Continuous

Measurement

10 Hours

Programmed

Sequence Time

24 Hours

Calibration Interval, recommended

24 Months

Battery

NiMH Batteries (5x C-cell), exchangeable

NiMH Batteries, built in

Dimensions, approximate

4.3 x 7.9 x 2.4 inches

(110 x 200 x 60 mm)

4.1 x 4.1 x 4.1 inches

(105 x 105 x 105 mm)

Weight, approximate

2.2 lbs. (1000 g)

2.2 lbs. (1000 g)

General Specifications