Electric and Magnetic Field Measurement

Specifications

magnetiC (b-) fielD

eleCtriC (e-) fielD

100 cm² Probe

internal Probe

3 cm Probe

Sensor System

Coil (internal or external)

Plate Electrode

Measurement Axis, selectable

Tri-Axial (Isotropic) or Single Axis

FIELD STRENGTH MODE

Frequency

Range

Broadband (+0/-3 dB),

selectable

5 Hz to 2 kHz, 30 Hz to 2 kHz, 5 Hz to 32 kHz or 30 Hz to 32 kHz

Band Pass / Band Reject

Filter, adjustable

15 Hz to 2 kHz (resolution 0.1 Hz)

Detection, selectable

RMS (averaging time 1 sec.)

Peak Value (proper phase)

Measurement

Range

Nominal

100 nT to 32 mT

100 nT to 32 mT

100 nT to 32 mT

10 V/m to 100 kV/m

Damage Level

(Peak)

Damage Level (Peak)

91 mT

a

@ ≤125 Hz

91 mT

a

@ ≤625 Hz

91 mT

a

@ ≤625 Hz

280 kV/m

Damage Level (Peak)

a

For magnetic field

probes depending on

frequency

Noise Level

(RSM), typical

Broadband,

30 Hz to 2 kHz

4 nT

100 nT

20 nT

0.7 V/m

Broadband,

5 Hz to 32 kHz

10 nT

200 nT

50 nT

4.5 V/m

Band Pass Filter,

50 Hz to 400 Hz

0.8 nT

25 nT

5 nT

0.14 V/m

Uncertainty,

typical

b

Broadband,

5 Hz to 2 kHz

±3% @ ≥40 nT

±5% @ ≥1µT

±4% @ ≥200 nT

±3% @ ≥5 V/m

Broadband,

5 Hz to 32 kHz

±3% @ ≥80 nT

±8% @ ≥2 µT

±5% @ ≥400 nT

±3% @ ≥40 V/m

Band Pass Filter, 50 Hz

to 400 Hz

±3% @ ≥10 nT

±5% @ ≥250 nT

±4% @ ≥50 nT

±3% @ ≥1 V/m

a The upper limit decreases linearly with increasing frequency above the mentioned frequency.

Overload limit for 100 cm² Probe =

Overload limit for 3 cm and internal Probe 100 cm² Probe =

b Uncertainty includes all partial uncertainties (absolute, linearity, frequency response, and isotropy) as well as temperature and humidity related deviations.

Signal sinusoidal, level >10% of selected measurement range; additional uncertainties apply with the steep frequency band limits.

efa-300 field analyzer

8000 mT • Hz

Frequency

( ) •

√ 2

40000 mT • Hz

Frequency

( ) •

√ 2